Characteristics and Specifications
9
Characteristics and Specifications
This section lists the characteristics and specifications for the probes. The probe
and oscilloscope should be warmed up for at least 20 minutes before any testing
and the environmental conditions should not exceed the probe’s specified limits.
Table 3 Electrical and Mechanical Characteristics
Description
Characteristic
Attenuation ratio
10:1
Bandwidth (–3 dB)
700 MHz
a
a.700 MHz BW only available on DSOX/MSOX 4000A-series oscilloscopes with 1 GHz or 1.5 GHz
bandwidth
Probe Risetime (10%-90%)
500 ps
Maximum Rated Input Voltage
400 V mains isolated
b
,
c
, 300 V CAT II
d
b.300 Vrms, 400 V (DC + peak AC) mains isolated; 0 V transient overvoltage with InfiniiVision 6000X
c.Maximum pulse rating = 1,650 V (step 0 V to 1,650 V), mains isolated, no overshoot permitted,
connected to oscilloscope with an input impedance of 1 M
Ω
± 1%, not applicable with
InfiniiVision 6000X.
d.Measurement Category II
Input Resistance (scope + probe)
10 M
Ω
Input Capacitance (system)
9.5 pF
Compensation Range
10 - 25 pF
Input Coupling of the Measuring Instru-
ment
1 M
Ω
Weight (probe only)
48 g
Cable Length
1.3 m
Probe Barrel Diameter
2.5 mm
Table 4 Safety Specifications
Specification
CEI/IEC 61010-031
N2894_97001.book Page 9 Monday, July 22, 2019 3:03 PM