20
N2795/6A User’s Guide
1
Using the Probes
Functional Test
Use the following generic measurement procedure to ensure that
your probe is functioning properly.
1
Connect the probe to an oscilloscope channel input and ensure
the input impedance of the oscilloscope matches the output
impedance of the probe (50
W
).
2
Connect a grounding accessory to the probe and connect the
ground to the ground terminal on the oscilloscope.
3
Connect the probe tip to the oscilloscope’s probe compensation
output.
4
Enable autoscale on the oscilloscope.
5
A square wave should now be displayed on the oscilloscope, if
the probe is functioning properly.