Calibrate the Oscilloscopes in the MultiScope System / 23
Use MultiScope Control Software / 24
Performing Channel Deskew / 26
Calibrate the Oscilloscopes in the MultiScope System
For best performance under any conditions, it is useful to verify the current
operating temperature of the oscilloscopes in the rack(s) or stackmount
configuration once the entire system has reached a stable thermal condition. The
temperature deviation from the previous calibration is displayed in the Calibration
dialog box of each oscilloscope.
If the inlet air temperature is not within the calibration temperature range, the
oscilloscope will provide a visual warning on the screen.
The oscilloscopes are warranted to meet all specifications when the ambient
temperature is ±5
°
C of the calibration temperature.