Step 4: Install the Modules
a. Install the module into the slot of the
chassis by placing the module card edges
into the front module guides (top and bot-
tom).
b. Slide the module to the rear of the chassis
and ensure that the injector/ejector handle
is pushed down in the unlatched (down-
ward) position.
c. Slide the module completely into the
chassis. When you begin to feel resistance,
pull up on the injector/ejector handle to
fully inject the module into the chassis.
7. Secure all modules to the chassis using the module mounting screws. Use a #1 Pozidriv® or a slotted screwdriver and
torque to 5 Lb-In (0.57 N-m). Performance may suffer if the screws are not tightened properly.
8. Verify that the PXI chassis fans are operable and free of dust and other contaminants that may restrict airflow.
9. Install filler panels and slot blockers after installing the modules. Missing filler panels or slot blockers may disrupt air cir-
culation in the chassis. The left-most slot does not accept a slot blocker.
10. Your shipment included an Keysight M9310A PXIe Source Output. With this module you received a straight, coaxial
adapter (1250-2316) (SMA female to SMA male). Connect this adapter to the RF Out connector of each M9310A, and
torque to 8 Lb-In (0.904 Nm). This will extend the life of this often-used connector.
11. Use the Cabling Diagram plus the Cable and Module Table on the next page to attach the cables to the instrument. The
torque specification for SMA connectors is 8 Lb-In (0.904 Nm).
12. If you are using a PCIe Cable Interface, such as the Keysight M9021, connect the Cable Interface in the chassis to the
PC host per the instructions that came with the Cable Interface.
13. Power up the PXI chassis.
14. If you are using a remote PC, reboot the PC host.
15. Proceed to
Step 5: Verify Operation of the Keysight M9380A PXIe CW Source (page 30)
.