Step 5: Verify Operation of the Keysight M9380A PXIe CW Source
Step 5: Verify Operation of the Keysight M9380A PXIe CW Source
In this step you will verify correct operation of the Keysight M9380A PXIe CW Source. Before running a Self
Test or performing a Calibration, assure that all required software is installed, the chassis is powered on,
and all cabling is correct.
See "M9380A Instrument Connections" on page 20
" for proper cabling.
The first step in this process is to conduct a Self Test of the M9380A.
1. Open the M9380A soft front panel (SFP) by selecting Start > All Programs > Keysight > M938x > M9380A SFP.
2. Upon opening the SFP, you are presented with the "Connect to Instrument" dialog. Select all the modules that are
components of the M9380A and press Connect. For example:
If no modules are visible in the SFP
"Connect to Instrument" dialog,
.
3. Conduct a Self Test (Utilities > Self Test... > Run Self Test). The M9380A Self Test runs each module through its own
Self Test.
If the Self Test passes (see results below), go to
Step 6: Installation is Complete (page 31)
.