Performance Verification
Performance verification ensures that the instrument performs within the specifications stated in the data sheet
(
http://literature.cdn.keysight.com/litweb/pdf/3120-1430EN.pdf
)
Recommended test equipment
The test equipments recommended for the performance verification and adjustment procedures are listed below. If
the exact instrument is not available, use the accuracy requirements shown to select substitute calibration
standards.
Type
Specification
Recommended model
Digital multimeter
Readout: 6 1/2 digits Basic DC Accuracy: 0.0035%
Keysight 34465A or equivalent
Current monitor
10 A (1 Ω), TC = 4 ppm/°C
100 A (0.01 Ω), TC = 4 ppm/°C
Guildline 9230A-10
Guildline 9230A-100
Power Supplies
150 V, 60 A, 350 W minimum
Keysight N7977A (160 V, 12.5 A) and
Keysight N7971A (20 V, 100 A)
Test considerations
– Ensure that the calibration ambient temperature is stable at 23 °C ± 5 °C.
– Ensure ambient relative humidity is less than 70%.
– Allow a 30 minutes warm-up period before verification or calibration.
– Keep cables as short as possible, consistent with the impedance requirements.
– Performance verification and calibration procedure can be performed either through front panel or rear panel
input.
The tests should be performed by qualified personnel. During performance verification tests, hazardous voltages
may be present at the outputs of the power supply.
14
Keysight EL30000 Series Service Guide