Keysight E1418A Service Guide
43
Performance Test Record
Verification Tests
DC Compliance Current Measurements
The 100 V range is used for the compliance current test. (The compliance current
is inferred from the voltage developed across the resistor).
M.U. = ± (0.035% of r 0.0006% of range)
M.U. = ± (0.00042 + 0.0006) = 0.00102 V
DC Short Circuit Current Measurements
The 100 mA range is used for the short circuit current measurement.
M.U. = ± (0.030% of r 0.005% of range)
M.U. = ± (0. 0.000005) = 0.000023 A
Current Output Measurements
The 100 mA range is used for the current output measurements.
M.U. = ± (0.030% of r 0.005% of range)
Test Accuracy Ratio (TAR)
The Test Accuracy Ratio (TAR) is defined as:
Where Maximum Allowed, Input, and Measurement Uncertainty are all in Vdc or
Amps.
The TAR is listed in table 4-1.
E1418A Output
% of reading
% of range
Measurement
Uncertainty
+ 0.020 A
0.000006 A
0.000005 A
± 0.000011 A
+ 0.015 A
0.0000045 A
0.000005 A
± 0.0000095 A
+ 0.010 A
0.000003 A
0.000005 A
± 0.000008 A
+ 0.005 A
0.0000015 A
0.000005 A
± 0.0000065 A
0.000 A
0.00000 A
0.000005 A
± 0.000005 A
– 0.005 A
0.0000015 A
0.000005 A
± 0.0000065 A
– 0.010 A
0.000003 A
0.000005 A
± 0.000008 A
– 0.015 A
0.0000045 A
0.000005 A
± 0.0000095 A
– 0.020 A
0.000006 A
0.000005 A
± 0.000011 A
TAR
MaximumAllowed Input
–
MeasurementUncertainty
----------------------------------------------------------------------------
=
Summary of Contents for E1418A Option 002
Page 1: ...Keysight E1418A 8 16 Channel D A Converter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...6 ...
Page 10: ...10 Keysight E1418A Service Guide ...
Page 32: ...26 Keysight E1418A Service Guide Operating Information Operator s Checks ...
Page 77: ...Keysight E1418A Service Guide 71 Adjustment Procedure Adjustments ...
Page 78: ...72 Keysight E1418A Service Guide Adjustments Adjustment Procedure ...
Page 90: ...84 Keysight E1418A Service Guide Service Repair Maintenance Guidelines ...
Page 100: ...94 Keysight E1418A Service Guide Replaceable Parts Ordering Information ...
Page 102: ...96 Keysight E1418A Service Guide Manual Changes Introduction ...
Page 105: ......