22
Keysight E1418A Service Guide
Operating Information
Preventive Maintenance
Cleaning Procedure
Use the following procedure to clean the Keysight E1418A.
Terminal Module
1
Remove the terminal module from the main module.
2
Disconnect all external wiring.
3
Clean the terminal module covers with a mild soap solution and lint-free
cloth.
4
Remove the terminal module cover and blow any dust or debris from the case.
5
Reassemble the terminal module.
6
Reconnect the external wiring.
Main Module
1
Remove the terminal module from the main module.
2
Clean the main module face plate with a mild soap solution and lint-free
cloth.
3
Disassemble the module following the instructions in Chapter 6.
4
Using an airgun with a grounded nozzle, remove dust from the PCA
assemblies.
5
Reassemble the module.
Disassembling the module exposes the assemblies to ESD damage
and is usually unnecessary. Perform steps 3 through 5 only if the
module is disassembled for repairs or reconfiguration.
Summary of Contents for E1418A Option 002
Page 1: ...Keysight E1418A 8 16 Channel D A Converter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...6 ...
Page 10: ...10 Keysight E1418A Service Guide ...
Page 32: ...26 Keysight E1418A Service Guide Operating Information Operator s Checks ...
Page 77: ...Keysight E1418A Service Guide 71 Adjustment Procedure Adjustments ...
Page 78: ...72 Keysight E1418A Service Guide Adjustments Adjustment Procedure ...
Page 90: ...84 Keysight E1418A Service Guide Service Repair Maintenance Guidelines ...
Page 100: ...94 Keysight E1418A Service Guide Replaceable Parts Ordering Information ...
Page 102: ...96 Keysight E1418A Service Guide Manual Changes Introduction ...
Page 105: ......