4
Installation and Verification
32
Keysight 8710x Operating and Service Manual
Quick check procedure
1
Connect the common port of the switch to Port 1 of the network analyzer and one of the outer
RF ports to Port 2 of the network analyzer as illustrated in
2
For standard drive, apply ground to the corresponding “drive” pin to close the selected path.
Refer to
3
For TTL drive (option T24), apply “High” to the corresponding “drive” pin to close the selected
path. Refer to
4
Perform the S-parameter measurement and verify against
“Supplemental specifications (cold
5
Repeat steps 1 to 4 until all paths are measured and verified.
Performance test
The coaxial multiport switches can be tested to the accuracy of the specifications with a network
analyzer or equivalent equipment of suitable accuracy. If a network analyzer is available, test the
instrument using the procedure in the analyzer’s operating manual.
Service instructions
Adjustment and repair
N1923/4A low PIM coaxial multiport switches do not require internal adjustments and are not
recommended for repair.
Maintenance
The connectors, particularly the connector faces, must be kept clean. For instructions on
connecting and care of your connectors, refer to the Microwave Connector Care Quick Reference
Card (08510-90360).
NOTE
If any of the low PIM coaxial multiport switches fails within the warranty period, a new unit will be
replaced. Refer to
on page 33 for more details.
Summary of Contents for 8710 Series
Page 1: ...Keysight 8710x Low PIM Coxial Multiport Switches Operating and Service Manual ...
Page 4: ...4 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 6: ...6 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 8: ...8 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 10: ...10 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...