Installation and Verification
4
Keysight 8710x Operating and Service Manual
31
Operating and Service Instructions
Operator’s check
The operator’s check is supplied to allow the operator to make a quick check on the coaxial
multiport switches prior to use or if a failure is suspected.
Description
The coaxial multiport switch is connected to a network analyzer configured for the S-parameter
measurement. The network analyzer may be set to sweep over the whole or selected frequency
range of the switch to be verified. The S-parameter measurement is the best way to determine if
the switch is working properly.
Figure 4-1
Connection to perform quick check
CAUTION
ESD exceeding the level specified in
or RF power applied is
greater than the maximum specified as in
may cause permanent
damage to the device.