Resolution
(continued)
•
Front-Panel Operation: Select either the slow or fast mode for each
resolution setting. The default mode is 5 digits slow.
5: RESOLUTION (MEAS MENU)
See also “To Set the Resolution,” on page 37.
•
Remote Interface Operation: You can set the resolution using the
following commands.
Specify the resolution in the same units as the measurement
function, not in number of digits. For example, for dc volts, specify the
resolution in volts. For frequency, specify the resolution in hertz.
CONF:VOLT:DC 10,0.001
4
1
⁄
2
digits on the 10 Vdc range
MEAS:CURR:AC? 1,1E-6
6
1
⁄
2
digits on the 1 A range
CONF:FREQ 1 KHZ,0.1 Hz
1000 Hz input, 0.1 Hz resolution
VOLT:AC:RES 0.05
50 mV resolution on the ac function
CONFigure:<
function
> {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
MEASure:<
function
>? {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
<
function
>:RESolution {<
resolution
>|MIN|MAX}
Chapter 4 Calibration Procedures
Understanding Resolution
88