See rear panel for proper fuse rating.
Agilent Part Number: 2110-0817 (250 mAT)
100, 120, 220 (230) or 240 Vac
1
Remove the power cord
. Remove the
fuse-holder assembly from the rear panel.
2
Remove the line-voltage selector from
the assembly.
3
Rotate the line-voltage selector until the
correct voltage appears in the window.
4
Replace the fuse-holder assembly in
the rear panel.
Verify that the correct line voltage is selected and the power-line fuse is good.