![Keysight 16861A Installation And Quick Start Manual Download Page 42](http://html1.mh-extra.com/html/keysight/16861a/16861a_installation-and-quick-start-manual_1961223042.webp)
42
Keysight 16860 Series Portable Logic Analyzers Installation/Quick Start Guide
4
Solving Problems
If there are touch screen problems
If you observe the following touch screen problems:
• If the Touch Off button on the instrument’s front panel doesn’t work.
• If there are problems with the right-click functionality on the touch
screen.
• If the accuracy of touch locations is off.
You may be able to calibrate the touch screen to fix the problem by
following the procedures described below for 16860 series of logic
analyzers.
1 From the Windows Start menu, choose
Start>All Programs>Keysight Logic
Analyzer>Utilities>Touch Screen>Touch Screen Calibration and Settings
.
2 In the Microchip TSHARC Control Panel dialog, select the
Calibration
tab.