Keysight R, Q, U, V, and W11644A 4-9
Use, Maintenance, and Care of the Devices
Handling and Storage
Handling and Storage
— Install the protective end caps and store the calibration devices in the
foam-lined storage case when not in use.
— Never store devices loose in a box, or in a desk or bench drawer. This is the
most common cause of device damage during storage.
— Keep devices clean.
— Do not touch mating plane surfaces. Natural skin oils and microscopic
particles of dirt are easily transferred to a device and are very difficult to
remove.
— Do not set devices contact-end down on a hard surface. The plating and the
mating plane surfaces can be damaged if the interface comes in contact
with any hard surface.