U3020AS16
Front and Rear Panel Features
Figure 3
Rear Panel
GPIB Connector
This connector allows the
Test Set
to be connected directly to a controller.
Factory set address = 26.
Refer to
Configuring the GPIB Interface
for instructions on how to change the GPIB address.
LAN
The instrument is controlled over Local Area Network (LAN).
Line Input
The line input contains the power cord receptacle.
Install the instrument so that the detachable power cord is readily identifiable
and is easily reached by the operator. The detachable power cord is the
instrument disconnecting device. It disconnects the mains circuits from the
mains supply before other parts of the instrument. The front panel switch is
only a standby switch and is not a LINE switch. Alternatively, an externally
installed switch or circuit breaker (which is readily identifiable and is easily
reached by the operator) may be used as a disconnecting device.
Always use the three-prong AC power cord supplied with this product. Failure
to ensure adequate grounding by not using this cord may cause damage to
the product.
GPIB
LAN Line Input
14
Keysight U3020AS16 User's and Service Guide