U3020AS16
General Performance
1-
General Performance
The Test Set performance is specified for a standalone instrument.
This product has an autoranging line voltage input. Be sure the supply
voltage is within the specified range.
Power Requirements
—
100-240V AC (50/60 Hz)
—
The instruments can operate with mains supply voltage fluctuations up to ±
10% of the nominal voltage.
—
Air conditioning equipment (or other motor–operated equipment) should
not be placed on the same
AC
line that powers the Test Set.
—
U3022AS
16
maximum power is 50 W.
Environmental Tests
The U3020AS16 complies with all applicable safety and regulatory
requirements for the intended location of use.
— Operating Environment (Indoor Use)
— Operating Temperature 0 to 40 °C
— Storage Temperature:
−
40 to +70 °C
— Operating Altitude: 0 to 2000 meters (~ 6,562 feet)
— The instrument can safely operate in a relative humidity of 80% for
temperatures to 31 degrees C, decreasing linearly to 50% relative humidity
at 40 degrees C.
8
Keysight U3020AS16 User's and Service Guide