1-
U3020AS16
Instrument Settings
Configuring the GPIB Interface
Programming access to the Test Set is also available through the instrument's
GPIB interface. The GPIB connector is located on the rear panel of the
instrument.
1.
Select the Agilent Control icon > Agilent Connection Expert. If a GPIB card
is installed in your computer, the GPIB interface was configured during
installation of the IO libraries and is displayed in the Explorer Pane of the
“Welcome” Screen.
Figure
12
Agilent Connection Expert
Adding Instruments to the GPIB Configuration
1.
Select Add Instrument on the tool bar.
2.
Select GPIB Interface > OK.
3.
Select U3020A’s Address in the GPIB address drop-down list > OK.
(Factory set address = 26)
Changing the GPIB Address
The U3020A GPIB address can only be changed programmatically. The GPIB
address command is:
SYSTem:COMMunicate:GPIB:ADDRess <address>
When the address is changed, the new GPIB address is not updated in the
Agilent Connection Expert if the instrument was previously configured.
1.
From the Agilent Connection Expert application window, highlight the
instrument that's address was changed and click Change Properties in the
Configurable Properties application window, change the address to the
new address setting and select OK.
24
Keysight U3020AS16 User's and Service Guide