U3020AD01
General Performance
Environmental Requirements
Refer to the Network Analyzer documentation for environmental requirements.
Environmental Tests
The U3020AD01 complies with all applicable safety and regulatory requirements for the
intended location of use.
•
Operating Environment (Indoor Use)
•
Operating Ambient: Temperature 0 to 40 °C
•
Operating Altitude: 0 to 2000 meters (~ 6,562 feet)
•
The instrument can safely operate in a relative humidity of 80% for temperatures to
31 degrees C, decreasing linearly to 50% relative humidity at 40 degrees C.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature within the
appropriate range. Air conditioning capacity must be consistent with the rating listed in the
PNA standard documentation.
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient temperature of the
PNA-X and Test Set to be maintained within ± 1 °C of the ambient temperature at calibration.
Dimensions and Space Requirements
Standard installation of the U3020AD01 and PNA includes configuration and installation on a
customer provided lab bench or table top of adequate size and strength.
Table 2
Test Set Physical Parameters
Item
Value
Height
8.8 cm (3.5 in)
Width
42.5 cm (16.73 in)
Depth
58 cm (23 in)
Weight
9 kg (19.8 lb)