U3020AD01
Instrument Settings
Configuring the GPIB Interface
Programming access to the Test Set is also available through the instrument's GPIB interface.
The GPIB connector is located on the rear panel of the instrument.
1. Select the Keysight Control icon > Keysight Connection Expert. If a GPIB card is installed in
your computer, the GPIB interface was configured during installation of the IO libraries and
is displayed in the Explorer Pane of the “Welcome” Screen.
Figure
13
Keysight Connection Expert
Adding Instruments to the GPIB Configuration
1. Select Add Instrument on the tool bar.
2. Select GPIB Interface > OK.
3. Select U3020A’s Address in the GPIB address drop-down list >OK.
(Factory set address = 9)
Changing the GPIB Address
The U3020A GPIB address can only be changed programmatically. The GPIB address command
is:
SYSTem:COMMunicate:GPIB:ADDRess <address>
When the address is changed, the new GPIB address is
not
updated in the Keysight Connection
Expert if the instrument was previously configured.
1. From the Keysight Connection Expert application window, highlight the instrument that's
address was changed and click Change Properties in the Configurable Properties application
window, change the address to the new address setting and select OK.
Keysight U3020AD01 User's and Service Guide
24