Figure
7
Connection Expert, Command Execution1
8. To the left is a Task Guide pane. Select "Send Commands to this Instrument." If you
don't see this pane, select: View > Task Guide. A command window should appear as
shown in
Figure
7
below.
9. To confirm that the Connection Expert can control the Test Set, type the following
command as shown in the Command field: ROUT:SEQ:TRIG ATT1_10 [Send Command]
The 10 dB LED indicator on the Test Set front panel should be ON.
10. To begin the testing, set the path attenuator to 0 dB and type the following command as
shown in the Command field: ROUT:SEQ:TRIG ATT1_00 [Send Command]
All LED indicators on the Test Set should be OFF.
U3020AD01
Operational Verification
13