Service and Maintenance
2
Keysight U1818A/B Operating and Service Manual
29
Figure 2-22
Steps to remove the browser from the ergonomic handle
Handling SMA probe heads
The U1818A/B probe amplifier can become damaged when used with the N5380A
or E2695A SMA probe heads. Use the Keysight N5380-64701 SMA Head Support
to prevent damage. Make sure to plug the probe amplifier into the SMA head
before installing the SMA Head Support and do not attempt to plug or unplug the
SMA head from the probe amplifier while it is in the SMA head support housing.
Figure 2-23
shows how to attach the SMA head support using two provided
screws.