Installation Note N9038-90039
17
Option EXM – External Mixing Upgrade
Front Frame Assembly Replacement
21.
Using a T-10 driver verify that all screws in the instrument that are accessible at this point are
properly torqued to 9 inch-lbs.
22.
Using a 5/16" open end wrench verify that all cables in the instrument that are accessible at
this point are properly torqued to 10 inch-lbs.
23.
Referring to
, reattach the W6 cable to the A23 Limiter assembly P2 and the W8
cable to the A8 Motherboard assembly J9.
Figure 10
Front Frame Cable Connections
24.
Carefully install the front frame assembly onto the instrument chassis, making sure that no
cables are pinched and that the O-ring gaskets around the input connectors remain in place.