42 Installation Note N5222-90133
Description of this Upgrade
Step 21. Perform Post-Upgrade Adjustments and Calibration
-
The adjustments listed here generate new correction constants. The analyzer
must have a backup of this new data in case any of the data becomes
corrupted.
To store the backup data, perform these steps:
— Navigate to the EEPROM Backup Utility, located at:
—
Windows 7 --
C:\Program Files (x86)\Keysight\Network
Analyzer\Service\eebackup.exe
— Windows 10 --
C:\Program Files\Keysight\Network
Analyzer\Service\eebackup.exe
— Run the program.
— Click Backup EEPROM.
— Click Backup TSMB Mem.
— Click Backup Synth Mem. (Applies to Version 7 Synthesizers Only)
— Click Exit when the program has finished.
Figure 24
EEPROM Backup Menu
Operator’s Check
Perform the Operator’s Check to check the basic functionality of the analyzer.
For instructions, refer to the “Tests & Adjustments” chapter of the Service
Guide.
If you experience difficulty with the basic functioning of the analyzer, contact