background image

Installation Note N5222-90131                                                                                                                                  33

Description of the Upgrade
Installation Procedure for the Upgrade

Step 20. Assemble the A72 (Port 3) and A73 (Port 4) Bias Tee 

Combiner Assemblies 

Refer to 

Figure 13

 for this step of the procedure. New parts are listed in 

Table 2 

on page 11

. Use a T-10 TORX driver to tighten all screws.

Figure 13 

Bias Tee Combiner Assembly (N5242-00051 (Ports 3 and 4), 5087-7403, & 

0515-1227)

Summary of Contents for N5221B

Page 1: ... Capability Upgrade Kit for Version 7 Synthesizers For Models With Direct Digital Synthesis Only To Upgrade PNA N5221B and N5222B Option 220 to Option 420 Upgrade Kit Order Number N5221BU 680 and N5222BU 680 Keysight Kit Number N5222 60131 ...

Page 2: ...to U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license set forth in the EULA represents the exclusive authority by which the U S government may use modify distribute or disclose the Software The EULA and the license set forth therein does not require or ...

Page 3: ...couplers two additional receiver couplers two additional source attenuators two additional receiver attenuators a splitter a modified front panel an additional cable guard LFE bias tee combiners PC assembly low frequency extension LFE new cables Refer to Overview of the Installation Procedure on page 15 This repair must be done at a service center or a self maintainer service center Refer to Getti...

Page 4: ...ailable on the Web at http www keysight com find assist If you do not have access to the Internet please contact your Keysight field engineer If You Have Problems With the Upgrade Kit Contents Keysight stands behind the quality of the upgrade kit contents If you have problems with any item in the kit refer to www keysight com and the Contact Keysight link In any correspondence or telephone convers...

Page 5: ... extremely sensitive components that can be ruined if mishandled Follow instructions carefully when making cable connections especially wire harness connections The person performing the work accepts responsibility for the full cost of the repair or replacement of damaged components IMPORTANT This document contains references to legacy and to a new A25 HMA26 5 Multiplier Amplifier assembly Your mo...

Page 6: ... License Key Redemption on page 6 Verify the License Contents Refer to the license message you received from Keysight If the model number serial number or option number do not match those on the license message you received from Keysight you will not be able to install the license key file If this is the case contact Keysight for assistance Refer to Getting Assistance from Keysight on page 4 Ensur...

Page 7: ... the newer HMA26 5 part number N5240 60101 installed you may discard these parts A22 splitter 5087 7139 W42 N5222 20009 W43 N5222 20007 W44 N5222 20008 Figure 1 If you have the legacy 5086 7765 HMA26 5 please discard the N5245 20195 semi rigid cables Refer to Figure 1 on page 7 The new N5240 60101 HMA26 5 has the splitter integrated into the assembly Refer to Figure 1 on page 7 Comparison of Legac...

Page 8: ... When the PDF of the Service Guide is displayed scroll through the Contents section bookmarks to locate the information needed Protecting Your Workspace from Electrostatic Discharge For information click on the Chapter 1 bookmark Electrostatic Discharge Protection in the PDF Service Guide1 ESD Equipment Required for the Installation 1 See Downloading the Online PNA Service Guide on page 8 Descript...

Page 9: ...river set to 6 in lbs 0 68 N m 1 N A T 20 TORX driver set to 21 in lbs 2 38 N m 1 N A 5 16 in 8 mm nutsetter or open end torque wrench set to 8 in lbs 0 9 N m 1 N A 3 16 in 5 mm nutsetter or open end torque wrench set to 6 in lbs 0 68 N m 1 N A 5 8 in 16 mm nutsetter or open end torque wrench set to 21 in lbs 2 38 N m 1 N A 1 in 25 4 mm torque wrench set to 72 in lbs 8 15 N m 1 N A 9 mm nutsetter ...

Page 10: ... the Upgrade Products affected N5221B and N5222B Option 220 with x2 DDS installed Installation to be performed by Keysight service center or personnel qualified by Keysight Estimated installation time 5 hours Estimated adjustment time 3 hours Estimated full instrument calibration time 7 5 hours ...

Page 11: ...brick 2 1 5087 7829 A26 Test port 3 receiver bridge 2 5087 7757 A27 Test port 4 receiver bridge A30 Test port 3 coupler 2 5087 7813 A31 Test port 4 coupler Dress panel overlay 4 Port Option 405 420 and 425 1 N5222 80036 A35 Test port 3 source attenuator 2 33321 60077 A36 Test port 4 source attenuator A43 Test port 3 receiver attenuator 2 33321 60078 A44 Test port 4 receiver attenuator A72 LFE Bias...

Page 12: ... connector for test set front plate hardware supplied with connector 12 1250 3805 Nuts flanged 2 5022 1087 Clamp SMP port 3 and port 4 bias combiners 2 5023 3299 Cable assembly coaxial A06 A06 205 mm long Connect to A70 LFE board 2 8120 5017 Cable assembly flat ribbon 2 8120 0982 Gel pad between each test coupler and the test set front plate 4 E4403 20033 Subpanel test set 4 port 1 N5221 00007 Dre...

Page 13: ...e A21 HMA26 5 to A22 splitter 1 N5222 20009 W43a RF cable A22 splitter to A23 mixer brick 1 N5222 20007 W44a RF cable A22 splitter to A24 mixer brick 1 N5222 20008 W52 RF cable A23 mixer brick R1 to A20 IF multiplexer P411 1 N5242 60021 W53 RF cable A23 mixer brick R2 to A20 IF multiplexer P412 1 N5242 60022 W55 RF cable A24 mixer brick D to A20 IF multiplexer P801 1 N5242 60024 W56 RF cable A24 m...

Page 14: ...1 8121 0982 Ribbon cable A19 test set motherboard J207 to A44 port 4 receiver attenuator 1 8121 0982 Ribbon cable A19 test set motherboard J213 to A24 mixer brick 2 1 N5242 60006 Ribbon cable A19 test set motherboard J202 to A35 port 3 source attenuator 1 N5242 60008 Ribbon cable A19 test set motherboard J203 to A36 port 4 source attenuator 1 N5242 60008 Cable DC 2 pin to R A SMP Port 3 bias combi...

Page 15: ...y Extension LFE Cables 8120 5014 x1 8120 5021 x1 N5240 60097 x1 and N5240 60098 From the IF Multiplexer IF MUX Board Step 9 Remove Some of the Mixer Brick MXB Cables Step 10 Remove Some Gray Cables From the A20 IF Multiplexer Board Step 12 Remove Some Bottom Side Semi Rigid Test Set Cables Step 11 Remove the A70 LFE Board Bracket Step 13 Remove the A23 Mixer Brick Assembly Step 14 Assemble the A24...

Page 16: ...onnect Connect and Route Low Frequency Extension LFE Cables 8120 5014 x2 and 8120 5017 x3 and the Other Ends of the Cables Connected to the IF Multiplexer IF MUX Board Step 34 Install the New A70 Low Frequency Extension LFE Board Step 35 Connect Reconnect A71 A74 Bias Tee Combiner s Cables to A70 Low Frequency Extension LFE Board and the Other Ends of the New Cables Connected to the IF Multiplier ...

Page 17: ...Key Redemption on page 6 Verify that the model number serial number and option number information on the license key match those of the instrument on which this upgrade will be installed If the model number serial number or option number do not match those on your license key you will not be able to install the option If this is the case contact Keysight for assistance before beginning the install...

Page 18: ...tions click the Chapter 7 bookmark Removing the Covers in the PDF Service Guide1 Step 4 Remove the Front Panel Assembly For instructions click the Chapter 7 bookmark Removing and Replacing the Front Panel Assembly in the PDF Service Guide1 Step 5 Remove the A19 Test Set Motherboard For instructions click the Chapter 7 bookmark Removing and Replacing the A23 test set motherboard in the PDF Service ...

Page 19: ...E Cables 8120 5014 x1 8120 5021 x1 N5240 60097 x1 and N5240 60098 From the IF Multiplexer IF MUX Board on page 21 Save 8120 5014 x2 and 8120 x1 for reuse Discard 8120 5021 2 Disconnect the W208 N5240 60112 Source 1 and W210 N5240 60113 LO Source cables from the LFE board as shown items and The other end of the W208 N5240 60112 and W210 N5240 60113 needs to be disconnected from the A11 direct digit...

Page 20: ...board i e MB IF MUX LFE TSMB ribbon cable to LFE board J1 item Save for reuse Refer to Figure 3 2 Refer to Figure 3 for the following a Remove the A70 LFE board s 0515 0665 screws x2 It will not be reused item b Remove the A70 LFE board from the standoffs item Figure 3 Remove the A75 LFE board N5240 60089 N5291 60005 and 0515 0665 IMPORTANT Save all mounting hardware for reuse And for later steps ...

Page 21: ...reviously in the process in Step 6 Disconnect A71 and A74 Bias Tee Combiner s LFE Cables From the A70 Low Frequency Extension LFE Board and the Ends of the LFE Cables Connected to the IF Multiplier IF MUX Board on page 19 Refer to Figure 4 Save 8120 5014 cable for reuse Discard 8120 5021 cable Figure 4 Disconnecting the gray cables on the IF MUX board 8120 5014 x1 and 8120 5021 x1 IMPORTANT Before...

Page 22: ...e Upgrade Step 9 Remove Some of the Mixer Brick MXB Cables Remove one end of some the mixer brick MXB gray cables item through Refer to Figure 5 These cables are not reused and can be discarded Figure 5 Remove the one end of the IF gray cables to the IF MUX board N5242 60025 and N5242 60026 ...

Page 23: ...ove Some Gray Cables From the A20 IF Multiplexer Board Remove the IF Multiplexer SPAM gray cables N5247 60023 and N5247 60024 to the A20 IF MUX board as indicated in Figure 6 on page 23 items and These cables are not reused and can be discarded Figure 6 Remove the IF MUX Gray Cables N5247 60023 and N5247 60024 ...

Page 24: ... Description of the Upgrade Installation Procedure for the Upgrade Step 11 Remove the A70 LFE Board Bracket Remove the LFE bracket as indicated item Refer to Figure 7 Figure 7 Remove the Screws x4 and LFE Bracket 0515 0372 N5240 00011 ...

Page 25: ...oupler to front panel Port 2 A74 bias T combiner W117 N5222 20071 A32 port 2 coupler to front panel Port 2 CPLR ARM Be careful not to damage the center pins of the semirigid cables Some flexing of the cables may be necessary but do not over bend them Leave the gray flexible cables the wire harnesses and the ribbon cables connected where possible Any that are removed should be labeled for reconnect...

Page 26: ...reinstalled W4 N5222 20040 W3 to A25 port 1 receiver coupler W10 N5222 20043 W9 to A24 port 2 receiver coupler W13 N5222 20005 A25 port 1 receiver coupler to A33 reference mixer switch W36 N5222 20032 REF 1 RCVR R1 IN to A33 reference mixer switch W35 N5222 20033 A33 reference mixer switch to front panel REF 1 SOURCE OUT W30 N5222 20091 qty 6 Front panel jumper Step 13 Remove the A23 Mixer Brick A...

Page 27: ...11 of this document Figure 8 A24 Mixer Brick Assembly 5087 7829 0515 1410 and 0515 0372 IMPORTANT The A22 splitter 5087 7139 and N5222 20007 N5222 20008 and N5222 20009 cables are only used with a legacy HMA26 5 p n 5087 7765 If your PNA has a new N5240 60101 assembly installed then set aside these parts as spares for use in other PNAs with the older HMA26 5 or discard If you are unclear which HMA...

Page 28: ...0022 N5242 60023 and N5242 60024 1 The A22 splitter 5087 7139 and N5222 20007 N5222 20008 and N5222 20009 cables are only used with a legacy HMA26 5 p n 5087 7765 If your PNA has a new N5240 60101 assembly installed then set aside these parts as spares for use in other PNAs with the older HMA26 5 installed or discard If you are unclear which HMA26 5 assembly your PNA has installed refer to Figure ...

Page 29: ...lation Procedure for the Upgrade Step 15 Install the A23 A24 Mixer Brick Brick Assembly Follow instruction 1 shown in Figure 9 New parts are listed in Table 2 on page 11 of this document Figure 9 Mixer Brick Brick and Receiver Coupler Assemblies Installation 0515 0430 ...

Page 30: ...ies Follow the instructions shown in Figure 10 New parts are listed in Table 2 on page 11 of this document Figure 10 A26 and A27 Receiver Bridge Assemblies 5087 7757 N5242 00006 and 0515 1410 Step 17 Install the A26 and A27 Receiver Bridge Assemblies Follow instruction 2 shown in Figure 9 New parts are listed in Table 2 on page 11 of this document bridge ...

Page 31: ...le A35 A36 Source and A43 A44 Receiver Attenuators Refer to Figure 11 for this step of the procedure New parts are listed in Table 2 on page 11 Use a T 10 TORX driver to tighten all screws Figure 11 Assemble the Receiver and Source Attenuators N5242 60008 x4 N5242 00007 x2 33321 60078 x4 8121 0982 x2 0515 0272 x4 ...

Page 32: ...for the Upgrade Step 19 Install A35 A36 Source and A43 A44 Receiver Attenuators Refer to Figure 12 for this step of the procedure New parts are listed in Table 2 on page 11 Use a T 10 TORX driver to tighten all screws Figure 12 Install the Receiver and Source Attenuators 0515 00430 x4 ...

Page 33: ...e Step 20 Assemble the A72 Port 3 and A73 Port 4 Bias Tee Combiner Assemblies Refer to Figure 13 for this step of the procedure New parts are listed in Table 2 on page 11 Use a T 10 TORX driver to tighten all screws Figure 13 Bias Tee Combiner Assembly N5242 00051 Ports 3 and 4 5087 7403 0515 1227 ...

Page 34: ...r this step of the procedure New parts are listed in Table 2 on page 11 Use a T 10 TORX driver to tighten all screws 1 Install the N5240 60098 x2 cable as shown Note the orientation of the cable item 2 Add 5023 3299 x2 clamps and 0515 2141 x2 clamp screws as shown item Torque to 6 in lbs Figure 14 Install cables and clamps to bias tee combiners N5240 60098 5023 3299 and 0515 2141 ...

Page 35: ...re for the Upgrade Refer to Figure 15 for this step of the procedure New parts are listed in Table 2 on page 11 Torque to 9 in lbs 3 Install the bias tee combiner assemblies using 0515 0372 x4 screws as shown Figure 15 Bias Tee Combiner Installation using 0515 0372 0515 1227 screws ...

Page 36: ...he PNA For instructions click the Chapter 7 bookmark Removing and Replacing the A29 A32 Test Port Couplers in the PDF Service Guide1 2 Using pliers remove the adhesive bumper on the A29 test port 1 coupler and on the A32 test port 2 coupler 3 Follow the two instructions shown in Figure 16 New parts are listed in Table 2 on page 11 of this document Figure 16 A29 A32 Test Port Coupler Assembly E4403...

Page 37: ...ort test set front plate of the PNA Keep the screws and LED boards for reuse later 2 Remove the bulkhead connectors nuts and washers from the 2 port front plate to reuse later 3 Remove the 2 port test set front plate from the test set deck Keep the screws for reuse later but dispose of the test set front plate 4 Follow the two instructions shown in Figure 17 Figure 17 LED Board Assemblies and Test...

Page 38: ...se and the new bulkhead connectors included in the kit for the remainder of this step 2 From the back side of the test set front plate insert a bulkhead connector into a hole in the plate 3 Install 1x washer and 1x nut Hand tighten nut and ensure bulkhead connector hexagon nut on the back side of test set front plate is aligned to the test set subpanel hexagon indent 4 Repeat previous two steps fo...

Page 39: ...131 39 Description of the Upgrade Installation Procedure for the Upgrade Step 25 Install the 4 Port Coupler Plate Assembly to the Deck Follow the four instructions shown in Figure 19 Figure 19 Coupler Plate Assembly Installation ...

Page 40: ...he DDS Assembly A Remove the DDS Cables For the semirigid cables follow the two instructions shown in Figure 20 These cables W1 N5240 20122 and W41 N5240 20121 must be saved they will be reinstalled Remove any gray cables connected to the DDS assembly and note their location They will be re connected in Step 29 Install the NEW A11 x4 Direct Digital Synthesizer DDS Assembly and Cables Figure 20 Rem...

Page 41: ...0131 41 Description of the Upgrade Installation Procedure for the Upgrade B Remove the DDS Assembly Follow the two instructions shown in Figure 21 The x2 DDS assembly will not be reused Figure 21 Remove the Old x2 DDS Assembly ...

Page 42: ...f the Upgrade Installation Procedure for the Upgrade Step 27 Assemble the A5 and A8 26 5 GHz Source 1 and 2 Assemblies Follow the six instructions shown in Figure 22 Figure 22 A5 Source 1 Assembly 5087 7342 N5222 20062 N5222 20063 and N5222 20064 1 ...

Page 43: ...131 43 Description of the Upgrade Installation Procedure for the Upgrade Step 28 Install the A8 26 5 GHz Source 2 Assembly Follow the three instructions shown in Figure 23 Figure 23 A8 Source 2 Assembly Installation 0515 0380 x2 ...

Page 44: ...DDS LO J5 2 Reconnect W1 N5240 20121 to DDS J4 and to Source 1 RF IN 3 Connect cable W2 N5240 20123 between the A8 source 2 board RF IN and the A11 source 2 J6 DDS assembly Tighten the cable connectors to 10 in lbs using a 5 16 in torque wrench reuse W41 N5240 20122 A11 13 5 GHz LO DD synthesizer assembly J5 to A21 HMA26 5 RF IN One end was left connected in A Remove the DDS Cables on page 40 reus...

Page 45: ...222 90131 45 Description of the Upgrade Installation Procedure for the Upgrade Figure 24 Install Semirigid Cables to DDS Assembly and Source 1 and Source 2 N5240 20121 N5240 20122 N5240 20123 Torque all cables to 10 in lbs ...

Page 46: ... A11 DD synthesizer assembly J10 to rear panel REF OUT W200 N5240 60115 reuse SRC DD synthesizer assembly J16 to SPAM J3 Figure 25 Reinstall Following Flexible Cables to the DDS Assembly Top 1400 0249 8120 5063 x2 and N5240 60115 4 Connect the following new and used cables in the order listed Refer to Figure 26 on page 47 W59 N5242 60012 reuse A20 IF multiplexer board P3 to A12 SPAM board J1 W149 ...

Page 47: ...ultiplexer board P603 to A12 SPAM board J5 W63 N5242 60016 reuse A20 IF multiplexer board P803 to A12 SPAM board J6 W200 N5240 600115 A11 SRC DD Synthesizer assembly J16 to SPAM J3 Figure 26 Connect Other End of Bottom Flexible IF Multiplexer and Top Flexible Gray Cables to SPAM Board N5242 60012 N5247 60023 N5242 60014 5247 60024 N5242 60016 and N5240 60115 2 Port is Shown But 4 Port is Similar ...

Page 48: ... and N5242 60015 to the A20 IF MUX board as indicated in Figure 27 items and Figure 27 Connect the IF MUX SPAM gray Low Frequency Extension LFE gray cables to A20 IF MUX N5242 60013 and N5242 60015 Step 31 Reinstall A70 Low Frequency Extension LFE Board Bracket 1 Reinstall the A70 LFE board bracket by reversing the steps in Step 11 Remove the A70 LFE Board Bracket on page 24 Torque the 0515 0372 s...

Page 49: ...ixer Brick MXB Cables and Route Cables Reconnect the other end of the new mixer brick MXB cables item and all of the IF multiplexer IF MUX gray cables items through Refer to Figure 28 Figure 28 Connect the other end of the new IF gray cables to the IF MUX board N5242 60019 N5242 60020 N5242 60022 N5242 60023 and N5242 60024 ...

Page 50: ...50 Installation Note N5222 90131 Description of the Upgrade Installation Procedure for the Upgrade Figure 29 Route the existing IF gray cables ...

Page 51: ...IF MUX LFE TSMB ribbon cable as shown on the IF MUX board items through You will connect the other ends of the IF gray cables later on the process Refer to Figure 30 Figure 30 Connecting and routing the gray cables on the IF MUX board 8120 5014 x2 and 8120 5017 x3 The 8120 5014 x1 cable connection that was connected at P604 is now connected on P404 The new 8120 5017 cable is used for the P604 conn...

Page 52: ... 32 Install the New Mixer Brick MXB Cables and Route Cables on page 49 Refer to Figure 31 Figure 31 Routing the reconnected MXB and IF MUX Gray Cables on the IF MUX board Step 34 Install the New A70 Low Frequency Extension LFE Board 1 Reinstall the A70 LFE board N5291 60001 by reversing the steps in Step 7 Remove the A75 Low Frequency Extension LFE Board on page 20 Torque the 0515 0665 screws to 9...

Page 53: ...wn in Figure 32 i e one end was installed in Figure 30 on page 51 2 Reconnect the W208 N5240 60112 Source 1 and W210 N5240 60113 LO Source cables and connect W209 N5240 60114 Source 2 to the LFE board as shown items through The other ends of these cables are connected to the A11 direct digital synthesizer DDS x4 in a later step Refer to Figure 32 and see also Figure 30 on page 51 Figure 32 Connect...

Page 54: ... and Port4 Connectors and Install Clamps onto the Ferrite Beads on page 68 Follow instructions carefully when making cable connections especially wire harness connections Incorrect connections can destroy components resulting in additional customer costs Be careful not to damage the center pins of the semi rigid cables Some flexing of the cables may be necessary but do not over bend them To avoid ...

Page 55: ... parts are listed in Table 2 on page 11 For W26 and W166 refer to Figure 34 on page 61 W26 N5222 20034 A32 port 2 coupler to front panel Port 2 CPLR ARM W166 N5222 20118 A31 port 4 test port coupler to A73 port 4 Bias combiner For W22 W80 W91 W165 and W168 refer to Figure 35 on page 62 W91 N5222 20022 Port 4 RCVR D IN to A44 port 4 receiver attenuator W165 N5222 20122 Port 4 CPLR THRU to A73 port ...

Page 56: ... receiver coupler to front panel REF 4 SOURCE OUT For W18 W76 W89 W160 and W161 refer to Figure 37 on page 64 W89 N5222 20021 Port 3 RCVR C IN to A43 port 3 receiver attenuator W162 N5222 20115 A29 port 1 test coupler to A71 port 1 Bias combiner W163 N5222 20121 Port 3 CPLR THRU to A72 port 3 Bias combiner W18 N5222 20015 A30 port 3 coupler to front panel Port 3 CPLR ARM W76 N5222 20019 A35 port 3...

Page 57: ...A26 port 3 receiver coupler W38 N5222 20024 REF 3 RCVR R3 IN to A24 mixer brick R3 Refer to Figure 40 on page 67 W37 N5222 20101 A33 reference mixer switch to A23 mixer brick R1 W35 reuse N5222 20033 A33 reference mixer switch to front panel REF 1 SOURCE OUT Refer to the note shown in Figure 33 for information on installing W36 part number N5222 20032 W36 reuse N5222 20032 REF 1 RCVR R1 IN to A33 ...

Page 58: ...at attach the reference mixer switch to the test set deck were loosened torque these screws now to 9 in lbs W10 reuse N5222 20043 W9 to A28 port 2 receiver coupler W8 N5222 20042 W7 to A27 port 4 receiver coupler W6 N5222 20041 W5 to A26 port 3 receiver coupler W4 reuse N5222 20040 W3 to A25 port 1 receiver coupler Install New Cable s A2 1 HMA26 5 to A23 A24 Mixer Bricks If your instrument has a n...

Page 59: ...26 5 N5240 60101 installed you can discard these cables W42 N5222 20009 A21 HMA26 5 to A22 splitter NOTE Tighten both screws on the splitter to 9 in lbs W44 N5222 20008 A22 splitter to A24 mixer brick W43 N5222 20007 A22 splitter to A23 mixer brick See also Verify the Model Version of HMA26 5 Installed on page 7 You will need to remove the cap that is installed on the HMA26 5 top connector before ...

Page 60: ...60 Installation Note N5222 90131 Description of the Upgrade Installation Procedure for the Upgrade Figure 33 1 Install semirigid cables Example Shown is For Units With a A22 Splitter Installed ...

Page 61: ...Installation Note N5222 90131 61 Description of the Upgrade Installation Procedure for the Upgrade Figure 34 Install N5222 20118 and Verify A74 port 2 bias combiner semirigid cables N5222 20034 ...

Page 62: ...N5222 90131 Description of the Upgrade Installation Procedure for the Upgrade Figure 35 Verify A74 port 2 and install A73 port 4 bias combiner semirigid cables N5222 20018 N5222 20020 N5222 20022 N5222 20116 and N5222 20122 ...

Page 63: ...te N5222 90131 63 Description of the Upgrade Installation Procedure for the Upgrade Figure 36 Verify A74 port 2 and install A73 port 4 bias combiner semirigid cables N5222 20001 N5222 20025 N5222 20026 and N5222 20027 ...

Page 64: ...N5222 90131 Description of the Upgrade Installation Procedure for the Upgrade Figure 37 Verify A71 port 1 and install A72 port 3 bias combiner semirigid cables N5222 20021 N5222 20115 N5222 20121 N5222 20015 and N5222 20019 ...

Page 65: ...nstallation Note N5222 90131 65 Description of the Upgrade Installation Procedure for the Upgrade Figure 38 Verify A71 port 1 N5222 20119 and Install A72 port 3 bias combiner semirigid cable N5222 20117 ...

Page 66: ...66 Installation Note N5222 90131 Description of the Upgrade Installation Procedure for the Upgrade Figure 39 Install A72 port 2 and A73 port 4 bias combiner semirigid cables N5222 20117 and N5222 20119 ...

Page 67: ...Installation Note N5222 90131 67 Description of the Upgrade Installation Procedure for the Upgrade Figure 40 Install N5222 20101 semirigid cable and clamps 1400 0249 x3 ...

Page 68: ...nd Install Clamps onto the Ferrite Beads 3 Reinstall the N5240 60097 x1 N5240 60098 x1 and install the NEW N5240 60098 x2 cable as shown Note the orientation of the cable item through Torque to 10 in lbs Refer to Figure 41 and Figure 42 on page 69 Figure 41 Verify N5240 60097 A74 Port 2 and connect new N5240 60098 A73 Port 4 Bias Tee Combiner to A70 LFE Board and Install Clamps on Ferrite Beads N5...

Page 69: ...ocedure for the Upgrade Figure 42 Connect New N5240 60098 x1 from the A72 Port 3 and Verify N5240 60098 x1 A71 Port 1 Bias Tee Combiners to the A70 LFE Board and Install Clamps on Ferrite Beads N5240 60098 x2 and 1400 1391 x2 Verify Port 1 cable is properly routed and secured ...

Page 70: ...llow the instruction shown in Figure 43 in this document Figure 43 Bulkhead Connections Front Panel Step 38 Reinstall the A19 Test Set Motherboard For instructions on reinstalling the board click the Chapter 7 bookmark Removing and Replacing the A19 test set motherboard in the PDF Service Guide1 1 See Downloading the Online PNA Service Guide on page 8 ...

Page 71: ... mixer brick 2 8121 0982 A19 test set motherboard J206 to A43 port 3 receiver attenuator 8121 0982 A19 test set motherboard J207 to A44 port 4 receiver attenuator N5242 60008 A19 test set motherboard J202 to A35 port 3 source attenuator N5242 60008 A19 test set motherboard J203 to A36 port 4 source attenuator For more A19 cable connections refer to Step 40 Install Reinstall the A71 74 Bias Tee com...

Page 72: ... page 73 See also the Chapter 6 bookmarks Bottom Ribbon Cables and Wire Harnesses 2 port Option 205 S N Prefixes 6021 in the PDF Service Guide1 New parts are listed in Table 2 on page 11 1 Connect Reconnect A71 and A74 gray DC cables to the test set motherboard TSMB as follows Reuse N5240 60091 A19 test set motherboard J541 to A71 port 1 bias T combiner N5240 60091 A19 test set motherboard J543 to...

Page 73: ...rt 4 and reinstall Port 1 and Port 2 bias Tee combiner s gray DC bias cables to the test set motherboard N5240 60091 Route Cables 2 Route bias cables as shown Separate cables as much as possible It is OK to cross the cables Avoid running cables parallel or next to each other Avoid tie wrapping to semirigid cables Refer to Figure 44 on page 73 ...

Page 74: ...pgrade Install clamps onto the ferrite beaded cables 3 Install clamps onto bias cables as shown Refer to Figure 45 and to Figure 46 on page 75 Figure 45 Install the clamps onto the port 1 bias cables N5240 60097 x4 and 1400 1334 x2 as shown Close up example of clamps installed on a bias cable ...

Page 75: ...Installation Note N5222 90131 75 Description of the Upgrade Installation Procedure for the Upgrade Figure 46 Install the clamps onto the port 2 bias cables N5240 60097 and 1400 1334 x2 as shown ...

Page 76: ...Figure 47 New parts are listed in Table 2 on page 11 1 The analyzer should be positioned on its left side fans facing upwards as shown 2 Reconnect the previously installed W208 N5240 60112 item and W210 N5240 60113 item cables to the A11 3 Connect the new W209 N5240 60114 item flexible cable as indicated in Figure 47 Figure 47 New test set cables Connect Verify the other end of the N5240 60112 N52...

Page 77: ... the 2 Port lower front panel label item 2 Remove the 10 screws save the screws for reuse from the 2 port dress panel and remove the dress panel item and respectively 3 Reassemble the front panel s lower dress panel assembly with the new 4 port front frame N5240 00009 by reversing the order of step 2 in the instructions previously followed Figure 48 Replacing the Front Panel s Lower Dress Panel an...

Page 78: ...48 on page 77 New parts are listed in Table 2 on page 11 1 Remove the protective backing from the new front panel overlay N5222 80036 2 Loosely place the overlay in the recess on the lower front panel 3 Placing two fingers at the middle press the overlay firmly onto the frame while sliding your fingers in opposite directions towards the ends of the overlay Repeat on all areas of the overlay Step 4...

Page 79: ...ttom side of the PNA carefully fold or push down the ribbon cables and wires so they can t be pinched between the hardware and the outer cover Ribbon cables and wires must never be positioned on top of hardware Step 48 Reinstall the Inner Cover For instructions click the Chapter 7 bookmark Removing the Covers in the PDF Service Guide1 Step 49 Reinstall the Outer Cover For instructions click the Ch...

Page 80: ...ing A keyboard and mouse must be connected to the network analyzer Option 220 License Removal Procedure 1 To start the Keysight License Manager press Start Keysight License Manager Keysight License Manager A Keysight License Manager dialog box will appear 2 Right click the on the desired option and click Delete 3 In the Keysight License Manager dialog box that appears press or click Yes to confirm...

Page 81: ...e Key Redemption on page 6 2 Copy the license file s from the email s to the root directory of the USB flash drive More than one license file may be copied to the USB flash drive 3 Insert the USB flash drive into the PNA s USB drive slot 4 Verifying and editing the license file For these steps refer to the example in Figure 49 on page 82 a Verify your USB flash drive is connected to a PC b Open yo...

Page 82: ...ive and remove the USB flash drive from your PC 5 Connect the USB flash drive to the PNA Within 5 seconds the PNA should display a small New licenses installed message Else load the license key file s manually move your license file s to C Program Files Agilent licensing It may take Keysight License Manager an extra 5 seconds to enable the licenses 6 Disconnect the USB flash drive from the PNA 7 O...

Page 83: ...unning Press Help About NA and verify that Option 420 is listed in the PNA application 3 After successful installation of all upgrades some features require some adjustments to ensure the instrument meets its specified performance Refer to the following Web site http na support keysight com pna If the option is still not enabled or if your old option s have been removed contact Keysight Technologi...

Page 84: ...the specified adjustments have been performed the analyzer should operate and phase lock over its entire frequency range EEPROM Backup The analyzer uses arrays of correction constants to enable the analyzer to produce accurate leveled source signals and receive clean test signals These constants are stored in non volatile EEPROM memory and in flash memory files The adjustments listed here generate...

Page 85: ...heck to check the basic functionality of the analyzer For instructions click the Chapter 3 bookmark Tests and Adjustments in the PDF Service Guide1 If you experience difficulty with the basic functioning of the analyzer contact Keysight Refer to Contacting Keysight on page 4 Calibration Although the analyzer functions its performance relative to its specifications has not been verified It is recom...

Page 86: ...p 54 Prepare the PNA for the User 1 If necessary reinstall front jumper cables 2 Install the cable guards pushing them over the front jumper cables until the cushioning material touches the front panel of the PNA 3 Install the dust caps on the test ports 4 Clean the analyzer as needed using a damp cloth ...

Page 87: ...Installation Note Xxxxx xxxxx 3 ...

Page 88: ...This information is subject to change without notice Keysight Technologies 2000 2020 Edition 1 September 2020 www keysight com ...

Reviews: