command:
syntax:
description:
parameters:
response:
C#:
example:
:PARameter:OPTical:SOURce:S2PFile
:PARameter:OPTical:SOURce:S2PFile<wsp>“<path string>”
Specifies the name of the s2p file to use for additional adaptor de-
embedding on the source side. In default mode only evaluated for O/E
and O/O measurements.
“<string>” the file name and path enclosed in double quotes.
<string> file name and path of the s2p file on the LCA system.
(property) OptSrcFile
:PAR:OPT:SOUR:S2PF “c:\temp\test1.s2p”
command:
syntax:
description:
parameters:
response:
C#:
example:
:PARameter:RECeiver:ATTenuation?
:PARameter:RECeiver:ATTenuation?
Retrieves the optical attenuation on the receiver path. In default mode
only evaluated for E/O and O/O measurements
none
<double> attenuation value, default unit is dB
(property) RcvAttOpt_dB
:PAR:REC:ATT? -> 3
command:
syntax:
description:
parameters:
response:
C#:
example:
:PARameter:RECeiver:ATTenuation
:PARameter:RECeiver:ATTenuation <wsp><
attenuation
>
Specifies the optical attenuation on the receiver path. In default mode only
evaluated for E/O and O/O measurements
attenuation
<double> attenuation value, default unit is dB
none
(property) RcvAttOpt_dB
:PAR:REC:ATT 2
command:
syntax:
description:
parameters:
response:
C#:
example:
:PARameter:RECeiver:PATH:LENGth?
:PARameter:RECeiver:PATH:LENGth?
Retrieves the geometrical length of the receiver path in m. In
default mode only evaluated for E/O and O/O measurements
none
<double> The path length value, default unit is meter.
(property) RcvLengthOpt_m
:PAR:REC:PATH:LENG? -> 0.3
command:
syntax:
description:
parameters:
response:
C#:
example:
:PARameter:RECeiver:PATH:LENGth
:PARameter:RECeiver:PATH:LENGth<wsp> <
length
>
Specifies the geometrical length of the receiver path in m. In
default mode only evaluated for E/O and O/O measurements
length
<double> path length value, default unit is meter.
none.
(property) RcvLengthOpt_m
:PAR:REC:PATH:LENG 0.45
command:
syntax:
description:
parameters:
response:
example:
:
PARameter:RECeiver:REFR:INDex?
:
PARameter:RECeiver:REFR:INDex ?
Retrieves the refractive index of the receiver path in dB. In default
mode only evaluated for E/O and O/O measurements.
none
<double> the refractive index value, unit is dB.
(property) RcvRefIdx
:PAR:REC:REFR:IND? -> 0
Remote Operation
Keysight N437x Series Lightwave Component Analyzer, Programmer’s Guide