Triggers a measurement on the LCA. If you call a continuous measurement
while another measurement is running, the original measurement is stopped
without starting a new measurement
<string> SINGle | CONTinuous
none
(method)
Measure()
:MEAS:STAR CONT
command:
:NWA:MODel?
syntax:
:NWA:MODel?
description:
The model number of the network analyzer
parameters:
none
response:
string
C#:
(property)
NWAModel
example:
:NWA:MOD? -> N5245A
command:
:NWA:PORT:NUMBer?
syntax:
:NWA:PORT:NUMBer?
description:
The number of network analyzer ports
parameters:
none
response:
integer
C#:
(property)
NumNWAPorts
example:
:NWA:PORT:NUMBer? -> 4
command:
*OPC?
syntax:
*OPC?
description:
Retrieves the operation complete state
parameters:
none
response:
<string> 0 | 1
C#:
(method)
OperationComplete()
example:
*OPC? -> 1
command:
syntax:
description:
parameters:
response:
C#:
example:
:PARameter:ADVAnced:MODE?
:PARameter:ADVAnced:MODE?
Returns 1 if advanced mode is enabled.
In advanced mode you
can force the LCA to switch the laser on or off, independent of
the measurement type. You also have additional optical- and RF-
path de-embedding possibilities, or can apply additional de-
embedding on the receiver and the source side, independent of
the measurement type.
none
<string> 0 | 1
(property)
Advanced
:PAR:ADVA:MODE? -> 0
command:
syntax:
description:
parameters:
response:
C#:
example:
:PARameter:ADVAnced:MODE
:PARameter:ADVAnced:MODE<wsp>ON | OFF | 1 | 0
Enables or disables advance mode, which allows changing some
default settings
. In advanced mode you can force the LCA to switch
the laser on or off, independent of the measurement type. You also
have additional optical- and RF-path de-embedding possibilities, or can
apply additional de-embedding on the receiver and the source side,
independent of the measurement type.
<string> ON | OFF | 1 | 0
none
(property)
Advanced
:PAR:ADVA:MODE ON
Remote Operation
50
Keysight N437x Series Lightwave Component Analyzer, Programmer’s Guide