command:
:PARameter:MODUlator:BIAS:MODE
syntax:
:PARameter:MODUlator:BIAS:MODE<wsp>CONT|EVER|ONCE
description:
Specify how often a modulator bias voltage optimization will be performed
parameters:
<string>
CONTinuous|EVERysweep|ONCE
response:
none
C#:
(property)
ModBiasOptimization
example:
:PAR:MODU:BIAS:MODE EVER
command:
:PARameter:OPTical:INPut:POWer:HIGH?
synstax:
:PARameter:OPTical:INPut:POWer:HIGH?
description:
Returns the state of the high power input property.
parameters:
none
response:
<string> 1 | 0, input power high true = 1, false = 0
C#:
(property) HighPower_Input
example:
:PAR:OPT:INP:POWer:HIGH? -> 0
command:
:PARameter:OPTical:INPut:POWer:HIGH
syntax:
:PARameter:OPTical:INPut:POWer:HIGH<wsp>ON|OFF|1|0
description:
Gets the state of the high power input property.
parameters:
<string> ON | 1 enables high power input, OFF | 0 disables high power input
response:
none
C#:
(property) HighPower_Input
example:
:PAR:OPT:INP:POWer:HIGH ON
command:
:PARameter:OPTical:OUTput:POWer?
synstax:
:PARameter:OPTical:OUTput:POWer?
description:
Returns the optical output power of the LCA in dBm.
parameters:
none
response:
<double> power value, the default unit is dBm.
C#:
(property)
OpticalPower_dBm
example:
:PAR:OPT:OUT:POWer -> -1
command:
:PARameter:OPTical:OUTput:POWer
synstax:
:PARameter:OPTical:OUTput:POWer<ws><
power
>
description:
Specify the optical output power of the LCA in dBm.
parameters:
power
<double>, power value in dBm
response:
none.
C#:
(property)
OpticalPower_dBm
example:
:PAR:OPT:OUT:POWer -1
command:
:PARameter:OPTical:PATH:DEEMbedding?
synstax:
:PARameter:OPTical:PATH:DEEMbedding?
description:
Returns whether the whole optical path de-embedding is set on or off.
parameters:
none
response:
<string> 1 = optical path de-embedding is enabled, 0 = disabled
C#:
(property) UseOpticalConnData
example:
:PAR:OPT:PATH:DEEM? -> 1
command:
:PARameter:OPTical:PATH:DEEMbedding
syntax:
:PARameter:OPTical:PATH:DEEMbedding<wsp>ON|1|OFF|0
description:
Switches the whole optical path de-embedding on or off.
parameters:
<string> ON | 1 = enabled optical path de-embedding, OFF | 0 = disable
response:
none
C#:
(property) UseOpticalConnData
example:
:PAR:OPT:PATH:DEEM? -> 1
Remote Operation
Keysight N437x Series Lightwave Component Analyzer, Programmer’s Guide
53