M9602A/M9603A front panel 

As shown in Figure 1, the M9602A/M9603A has High force, Low force, High sense, 
Low sense and Guard. It supports both two-wire and four-wire measurement, and low 
current measurement with a guard. 

 

Step 2-1. Select cable and accessories for connection between M9602A/M9603A and your DUT 

You will need to select an appropriate accessory according to the connection type. 
There are basically three types of connection configurations. The essential cables and 
accessories are as follows: 

Two-wire or 
Four-wire 

Pulse width 

Recommended cable

 

High force 

High sense 

Low sense 

Four-wire 

<= 50 

μs

 

Triaxial(m) to BNC(f) 
+ Low inductance cable

1

 

High current triaxial cable

2

 

Triaxial-to-SMB cable

3

 

> 50 

μs or DC

 

High current triaxial cable 

High current triaxial cable 

Triaxial-to-SMB cable 

Two-wire 

 

High current triaxial cable 

NA 

NA 

1.  A low inductance cable does not have Guard, and is limited in low current measurement. Guard connection is 

required for low current measurement (~ 10 nA or less measurement.)   

2.  A low inductance cable can be used, though it is limited in low current measu rement. 
3.  An SMB-BNC cable can be used, however, do not connect the outer shield, Low Force, to the chassis/frame ground 

of the DUT when M9603A is grounded.     

 

 

 

 

 

 

 

 

 

 

 

 

 
Figure 1. M9602A/M9603A Front panel 

Two external trigger lines 

  

PXIe trigger lines are available on the 
backplane 

I

nterlock

 

 

One SMB terminal for Low sense 

Two triaxial terminals for high force / sense 

Low terminals

 

-

 

floating 

 

-

 

grounded 

 attach a short bar

 

 

 

High force

 

Guard

 

High sense

 

Low force

 

Summary of Contents for M9602A

Page 1: ...ptions and associated accessories This guide provides step by step instructions to help you configure the product to meet the desired test requirements Model number M9602A PXIe SMU 15 MSa s 1 pA 60 V...

Page 2: ...hort bar 1ea Short bar to connect low terminals to chassis common orderable part number is M9601 87001 includes 5 pcs Connector terminal block 2 5 mm 6 terminal 1ea Connector terminal block to connect...

Page 3: ...1 High current triaxial cable2 Triaxial to SMB cable3 50 s or DC High current triaxial cable High current triaxial cable Triaxial to SMB cable Two wire High current triaxial cable NA NA 1 A low induct...

Page 4: ...connection is required for low current measurement 10 nA or less measurement Do not connect the outer shield or Low Force to the chassis frame ground of the DUT when M9603A is grounded Figure 2 Connec...

Page 5: ...NC f adaptor instead PX0105A 001 002 Low inductance BNC cable 1 5 m 3 m 1 High force N1254A 106 Triaxial m to BNC f adaptor 1 Used with PX0105A Four wire connection for pulse 50 s or DC Figure 3 shows...

Page 6: ...e 4 shows the connection for 2 wire connection For this configuration select the items shown in Table 3 Table 3 Required items for 2 wire connection Product number Description Qty Note PX0104A 001 002...

Page 7: ...X0101A 001 or 002 BNC to ferrule terminal cable with a connector terminal block to connect the M9602A M9603A s interlock pins to it For more detailed information please refer to the Keysight M9602A M9...

Page 8: ...tion of calibration comes as standard The optional ISO 17025 not accredited ANSI Z540 and commercial calibration certificate with test data are available as follows Description Product number Addition...

Page 9: ...Connection between M9602A M9603A and your DUT Two wire connection or four wire connection If you are measuring very small resistances or applying a very large current you should use the four wire mea...

Page 10: ...urements 1 nA requires a guarding to prevent leakage through the measurement cable Figure 7 illustrates a simplified overview of the guarding technique Guarded measurements require the use of triaxial...

Reviews: