Getting Started
STEP 3. Install the Network Analyzer Modules
Figure 1-7
Case 2: Up to 12-port Module in Two Chassis Configuration
NOTE
In these instructions, four network analyzer modules are configured. Use these
instructions as a general guide when installing a different number of modules.
Direct Access Receiver (M9377A) Installation
If you are installing the direct access receiver, M9377A only (without the
directional coupler, M9378A), the installation procedure is the same as
standard receiver M9376A installation
Configurable Test Set (M9377A and M9378A/B) Installation
a. The installation of the Configurable Test Set differs from installation of
the Standard Receiver. Due to the semi rigid cables usage, there is
limitation of how 2-way and 4-way divider can be positioned. It is not
possible to use 4-way divider only for the Configurable Test Set
configuration. Refer to
“2-Way Divider and 4-Way Divider Installation”
to learn more on how to install the dividers on the modules.