:POD Commands
25
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
587
:POD<n>:THReshold
Command Syntax
:POD<n>:THReshold <type>[<suffix>]
<n> ::= An integer, 1 or 2, is attached as a suffix to the command and
defines the group of channels that are affected by the command.
<type> ::= {CMOS | ECL | TTL | <user defined value>}
<user defined value> ::= -8.00 to +8.00 in NR3 format
<suffix> ::= {V | mV | uV}
POD1 ::= D0-D7
POD2 ::= D8-D15
TTL ::= 1.4V
CMOS ::= 2.5V
ECL ::= -1.3V
The :POD<n>:THReshold command sets the threshold for the specified group of
channels. The threshold is used for triggering purposes and for displaying the
digital data as high (above the threshold) or low (below the threshold).
Query Syntax
:POD<n>:THReshold?
The :POD<n>:THReshold? query returns the threshold value for the specified group
of channels.
Return Format
<threshold><NL>
<threshold> ::= Floating point number in NR3 format
See Also
•
"Introduction to :POD<n> Commands"
•
•
Example Code
' THRESHOLD - This command is used to set the voltage threshold for
' the waveforms.
There are three preset values (TTL, CMOS, and ECL)
' and you can also set a user-defined threshold value between
' -8.0 volts and +8.0 volts.
'
' In this example, we set channels 0-7 to CMOS, then set channels
' 8-15 to a user-defined 2.0 volts, and then set the external trigger
' to TTL.
Of course, you only need to set the thresholds for the
' channels you will be using in your program.
NOTE
This command is only valid for the MSO models.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...