27
Keysight UXR-Series and MXR/EXR-Series MultiScope
Hardware Configuration Guide
5 Next Steps
Individually Calibrate the Oscilloscopes in the MultiScope System / 27
UXR-Series MultiScope Calibration Connections / 28
Use MultiScope Control Software / 30
Go To “Using Hosted Infiniium Oscilloscopes” in the Online Help / 31
With UXR-Series MultiScope systems, you make a long calibration cable and later
follow prompts by the Infiniium oscilloscope software to connect the cable to
individual oscilloscopes in the system as necessary.
With MXR/EXR-Series MultiScope systems, all the physical connections for
calibration have already been made (as described in the previous chapters of this
manual). In this case, you can let the Infiniium oscilloscope software perform
calibration without being prompted to change connections.
Individually Calibrate the Oscilloscopes in the MultiScope System
For best performance under any conditions, it is useful to verify the current
operating temperature of the oscilloscopes in the rack(s) or stackmount
configuration once the entire system has reached a stable thermal condition. The
temperature deviation from the previous calibration is displayed in the Calibration
dialog box of each oscilloscope.
If the inlet air temperature is not within the calibration temperature range, the
oscilloscope will provide a visual warning on the screen.
The oscilloscopes are warranted to meet all specifications when the ambient
temperature is ±5
°
C of the calibration temperature.