Two-Oscilloscope System, MXR/EXR-Series
4
Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide
25
Connections
Connections for Synchronization
The following two connections are required for synchronization.
1
Connect
TRIG OUT
from the Leader to
AUX TRIG IN
on the Follower through a
BNC cable.
2
Connect
10 MHz REF OUT
from the Leader to
10 MHz REF IN
on the Follower
through a BNC cable.
Figure 6
Connections for synchronization and reference clock skew calibration
10 MHz
REF OUT
AUX
OUT
TRIG
OUT
GEN
OUT
10 MHz
REF IN
AUX
TRIG IN
Leader
Front
Side
10 MHz
REF OUT
AUX
OUT
TRIG
OUT
GEN
OUT
10 MHz
REF IN
AUX
TRIG IN
Follower