Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide
2
Two-Oscilloscope System, UXR-Series
Connecting the Oscilloscopes
Using the parts in a N2164A (two in rack) expansion kit, follow these instructions:
1
Mount the 54964-60401 power divider assembly in the designated rack
location (with connectors facing back). See
2
Connect the four SMA 50
Ω
terminators (1250-3745) to the four unused
left-most power divider assembly outputs.
3
Connect one 54964-61632 cable from the Leader
Sample_CLK_Out
output to the
power divider input.
4
Connect two 54964-61632 cables between the power divider assembly outputs
and the oscilloscopes'
Sample_CLK_In
inputs.
5
Connect one 54964-61632 cable between the Leader
Trig Out
and Follower #1
Aux Trig In
connectors.
6
Connect one 54964-61632 cable between the Leader
Reference CLK Out
and
Follower #1
Reference CLK In
connectors.
Leader
Follower #1
Power divider
Place 1250-3745 terminators
on all unused divider outputs
Reference
CLK
In Out
Trig Out
Aux Trig
In
Sample
CLK
In Out
10 MHz
Out In
Connections
(Back)
Reference
CLK
In Out
Trig Out
Aux Trig
In
Sample
CLK
In Out
10 MHz
Out In