Keysight UXR-Series and 

MXR/EXR-Series MultiScope 

Hardware 

Configuration 

Guide

Summary of Contents for EXR Series

Page 1: ...Keysight UXR Series and MXR EXR Series MultiScope Hardware Configuration Guide...

Page 2: ...nd at www keysight com find sweula The license set forth in the EULA represents the exclusive authority by which the U S government may use modify distribute or disclose the Software The EULA and the...

Page 3: ...ssories containing the necessary cables adapters and power dividers are required N2164A UXR Series Multi Frame Base Kit First Two Frames N2165A UXR Series Multi Frame Expansion Kit Add One Frame The N...

Page 4: ...n an external host computer MultiScope software is included as a standard feature This software allows time synchronized acquisitions for multiple oscilloscope systems with very low oscilloscope to os...

Page 5: ...Included MultiScope systems also require Equipment rack s Rackmount kits for the oscilloscopes MXR2RACK for MXR Series or EXR2RACK for EXR Series MultiScope control software Remote interface connecti...

Page 6: ...the MultiScope system see Chapter 5 Next Steps starting on page 27 See Also The Infiniium UXR Series oscilloscope software s online help for information on using the MultiScope Hosted Instruments soft...

Page 7: ...es 14 3 Three to Five Oscilloscope Systems UXR Series Expansion Equipment Required 18 Mounting Oscilloscopes in Rack 19 Connecting the Equipment 20 Parts in the N2165A Add One Expansion Kit 20 Connect...

Page 8: ...es MultiScope Hardware Configuration Guide Infiniium Oscilloscope Software on the Leader Oscilloscope 30 D9010BSEO Infiniium Offline Oscilloscope Analysis Software 30 Go To Using Hosted Infiniium Osci...

Page 9: ...must be present to ensure that the oscilloscope delivers the best possible measurement accuracy For UXR Series MultiScope systems refer to the Infiniium UXR Series Real Time Oscilloscopes User s Guid...

Page 10: ...10 Keysight UXR Series and MXR EXR Series MultiScope Hardware Configuration Guide 1 MultiScope System Requirements...

Page 11: ...t 13 Expansion Equipment Required Table 1 Equipment Required for Two Oscilloscope Rack Mount System Model Description Quantity N2164A UXR Series Multi Frame Base Kit First Two Frames 1 N2156A Rack mou...

Page 12: ...uctions in the Rack Mount Kit User s Guide mount oscilloscopes in their designated rack locations NOTE Locate oscilloscopes in the rack so that the overall rack center of gravity is about half way up...

Page 13: ...ultiScope system These parts are used to create a long cable for performing channel deskew Part number Qty Description 54964 61632 5 1 m phase stable cable 54964 60401 1 6 way splitter assembly with r...

Page 14: ...uts 3 Connect one 54964 61632 cable from the Leader Sample_CLK_Out output to the power divider input 4 Connect two 54964 61632 cables between the power divider assembly outputs and the oscilloscopes S...

Page 15: ...tiScope Hardware Configuration Guide 15 7 Connect the oscilloscopes to the Leader oscilloscope or external host computer through the desired remote interface connection LAN USB etc 8 If these are the...

Page 16: ...16 Keysight UXR Series and MXR EXR Series MultiScope Hardware Configuration Guide 2 Two Oscilloscope System UXR Series...

Page 17: ...UXR Series and MXR EXR Series MultiScope Hardware Configuration Guide 3 Three to Five Oscilloscope Systems UXR Series Expansion Equipment Required 18 Mounting Oscilloscopes in Rack 19 Connecting the E...

Page 18: ...on Kit Add One Frame of scopes 2 N2156A Rack mount kit of scopes Figure 4 Expansion Accessories Required for Three to Five Oscilloscope Rack Mount System UXR Series oscilloscope UXR Series oscilloscop...

Page 19: ...s Guide mount oscilloscopes in their designated rack locations NOTE Locate oscilloscopes in the rack so that the overall rack center of gravity is about half way up the rack Because of cable lengths u...

Page 20: ...EXR Series MultiScope Hardware Configuration Guide 3 Three to Five Oscilloscope Systems UXR Series Connecting the Equipment Parts in the N2165A Add One Expansion Kit Part number Qty Description 54964...

Page 21: ...ider Place 1250 3745 terminator on all unused divider outputs Reference CLK In Out Trig Out Aux Trig In Sample CLK In Out 10 MHz Out In Reference CLK In Out Trig Out Aux Trig In Sample CLK In Out 10 M...

Page 22: ...3745 from the power divider output that will be used for the oscilloscope 2 Connect one 54964 61632 cable between the power divider output and the oscilloscope s Sample_CLK_In input 3 Connect one 549...

Page 23: ...Guide 4 Two Oscilloscope System MXR EXR Series MXR EXR Series Oscilloscope Requirements 24 Connections 25 Connections for Reference Clock Skew Calibration 26 This chapter describes how to configure t...

Page 24: ...scilloscope MultiScope systems made up of MXR EXR Series oscilloscopes you can use the N2124A multi frame cable kit Refer to your analysis software documentation for additional Infiniium application s...

Page 25: ...zation 1 Connect TRIG OUT from the Leader to AUX TRIG IN on the Follower through a BNC cable 2 Connect 10 MHz REF OUT from the Leader to 10 MHz REF IN on the Follower through a BNC cable Figure 6 Conn...

Page 26: ...for the optional drift correction 1 Connect the Follower s AUX OUT signal to the channel 4 inputs on both oscilloscopes using BNC cables and a passive power splitter or divider From the N2124A multi f...

Page 27: ...in the previous chapters of this manual In this case you can let the Infiniium oscilloscope software perform calibration without being prompted to change connections Individually Calibrate the Oscill...

Page 28: ...ope calibration is included in the N2164A UXR Series Multi Frame Base Kit These parts are used to create a long calibration cable The long calibration cable is connected from the Follower 1 oscillosco...

Page 29: ...061 5311 adapter 3 5 mm female to female 1 85 mm 5061 5311 adapter 3 5 mm female to female 1 0 mm 5067 1393 adapter 1 mm ruggedized female to 2 92 mm female Cal Out Calibration Cable Connections Follo...

Page 30: ...iniiScan From the Infiniium Offline software you have only edge and glitch trigger modes No additional computer is required to run the Infiniium Offline software No space for an additional computer is...

Page 31: ...finiium Oscilloscopes in the Online Help For instructions on Identifying oscilloscopes in the MultiScope system making software connections Time correlating oscilloscopes calibration Using the MultiSc...

Page 32: ...32 Keysight UXR Series and MXR EXR Series MultiScope Hardware Configuration Guide 5 Next Steps...

Page 33: ...ck mount 13 expansion accessories 3 F final steps 27 follower oscilloscope 3 5 H host computer 4 5 host computer requirements 30 I in the box MultiScope control software 30 IO Libraries Suite 30 M Mul...

Page 34: ...34 Keysight UXR Series and MXR EXR Series MultiScope Hardware Configuration Guide Index...