Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
TTV
The TTV command performs the high speed spot measurement, and returns the
measurement data and the time data for the time from when the timer is cleared until
the measurement is started. The command starts a voltage measurement regardless
of the SMU operation mode, trigger mode (TM command), and measurement mode
(MM command).
This command is effective for ASCII data output format. Refer to
.
Execution
Conditions
CN command has been executed for the specified channel.
Syntax
TTV
chnum
[,
range
]
Parameters
chnum
:
Measurement channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
.
range
:
Measurement range or ranging type. Integer expression. If you select
the fixed range, the instrument performs measurement by using the
specified range. For the auto or limited auto ranging, the measurement
range will be set to the minimum range that covers the measured
values. However, the instrument never uses the range less than the
specified range for the limited auto ranging. See
.
If you do not specify the
range
parameter for current output channels,
the channel uses the minimum range that covers the compliance value.
The
range
parameter is meaningless for voltage output channels. The
measurement ranging type is always same as the output ranging type.
Remarks
To read the time data with the best resolution (100
μ
s), the timer must be cleared
every 100 sec or less for the FMT1, 2, or 5 data output format, or
every 1000 sec or less for the FMT 11, 12, 15, 21, 22, or 25 data output format.
Example
Statements
OUTPUT @E5270;"TTV 1"
ENTER @E5270 USING "#,5X,13D,X";Time
ENTER @E5270 USING "#,5X,13D,X";Vdata
PRINT "Data=";Vdata*1000;"mV, at";Time;"s"
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......