Keysight CX3300 User’s Guide, Edition 5
99
Performing User Calibration
Using the CX1105A
Using the CX1105A
The CX3300 provides the offset user calibration for the CX1105A.
The user calibration data, Offset [V], is acquired and is used for error correction.
Effective for reducing the offset error.
Preparation
Before starting the user calibration, connect the CX1105A to the CX3300 analog
input channel.
Prepare the test adapter.
After the connection, warm up them for 30 minutes.
Procedure
You can perform the user calibration as shown below.
1. Press the
Menu
key several times to open the Configuration dialog box. The
Menu
key is located above the Horizontal control on the front panel.
2. Click
User Calibration
to display the Configuration > User Calibration screen.
3. Click
Execute
for the channel under user calibration.
You will see a dialog box which asks you to connect the test adapter to the
CX1105A input. Then connect them as shown in
and click Yes.
The user calibration will start.