Keysight CX3300 User’s Guide, Edition 5
25
Setting Up the CX3300
Verifying Basic Operation
The 2 mA DC current is applied from the Aux Out terminal.
5. Click the close (x) button located in the upper right corner on the Configuration
dialog box to close the dialog box.
6. Press the
Run
key to start waveform acquisition.
7. Use the Vertical control knob to adjust the vertical scale. You will see DC
current as shown in
Figure 1-4
Calibration Output
If You Use the CX1104A
Required accessories
•
Extension cable, 1 ea., furnished with the CX1104A
•
Resistive sensor head (CX1211A, CX1212A, CX1213A, CX1214A, CX1215A, or
CX1216A), 1 ea.,
must be prepared by user
•
Banana adapter, 1 ea., furnished with the CX1104A
•
Banana adapter (Banana plug to BNC jack), 1 ea.,
must be prepared by user
•
Coaxial cable (BNC plug to BNC plug), 1 ea., furnished with the CX3300
Procedure
1. Connect the CX1104A to the Aux Out terminal on the CX3300 front panel by
using the accessories listed above.
Do not forget to fasten the screws on the sensor head to assure the contact
with the furnished banana adapter.