Keysight CX3300 User’s Guide, Edition 5
89
Performing User Calibration
Using the CX1101A/CX1102A/CX1103A
Figure 3-1
Connection Diagram for User Calibration (CX1101A/CX1102A/CX1103A)
After the channel passed the user calibration, the ON/OFF switch will be available.
•
Set ON to enable error correction.
Error correction is performed for the data measured by this channel from now
on.
•
Set OFF to disable error correction.
Error correction is not performed.
If the channel failed the user calibration, the ON/OFF switch is not available. Error
correction is not performed.
Once you disconnect the CX1101A/CX1102A/CX1103A from the CX3300, the user
calibration data for this channel will be cleared.
CX1101A/CX1102A with CX1203A
or
CX1103A
SMA-to-BNC adapter
(furnished with
current sensor)
BNC cable
(furnished with CX3300)
CX3300