85058V
1-5
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the gender of cable
connectors and adapters is referred to in terms of the center conductor. For example, a
connector or device designated as 1.85 mm –f– has a 1.85 mm female center conductor.
A connector gage is referred to in terms of the connector that it measures. For instance, a male
connector gage has a female connector on the gage so that it can measure male devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit include:
• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques
All of these are described in
. Failure to detect and remove dirt or metallic particles on
a mating plane surface can degrade repeatability and accuracy and can damage any connector
mated to it. Improper connections, resulting from pin depth values being out of specification
(see
), or from bad connection techniques, can also damage these
devices.
Summary of Contents for 85058V
Page 1: ...Keysight Technologies 85058V 1 85 mm Verification Kit User s and Service Guide...
Page 2: ......
Page 3: ...85058V...
Page 7: ...1 General Information...
Page 12: ...1 6 85058V General Information Clarifying the Terminology of a Connector Interface...
Page 13: ...2 Specifications...
Page 20: ...2 8 85058V Specifications Electrical Specifications...
Page 21: ...3 Use Maintenance and Care of the Devices...
Page 36: ...3 16 85058V Use Maintenance and Care of the Devices Handling and Storage...
Page 37: ...4 Performance Verification...
Page 40: ...4 4 85058V Performance Verification Recertification...
Page 41: ...5 Troubleshooting...
Page 43: ...85058V 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart...
Page 47: ...6 Replaceable Parts...
Page 54: ...6 8 85058V Replaceable Parts Replaceable Parts...