85058V
3-9
Use, Maintenance, and Care of the Devices
Gaging Connectors
NOTE
Always hold a connector gage by the gage barrel, below the dial indicator. This
gives the best stability, and improves measurement accuracy. (Cradling the gage
in your hand or holding it by the dial applies stress to the gage plunger
mechanism through the dial indicator housing.)
1. Select the proper gage for your connector. The 1.85 mm connectors are gaged with the
same gages and in the same way as precision 2.4 mm connectors. (Refer to
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to
“Cleaning Connectors” on page 3-6
3. Zero the connector gage (refer to
):
a. While holding the gage by the barrel, and without turning the gage or the gage master,
screw the gage master connecting nut onto the male gage, just until you meet
resistance. Connect the nut finger tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting nut to
90 N-cm (8 in-lb). Refer to
c. Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer
corresponds to the correction value noted on the gage master. Do not adjust the gage
dial to zero, unless the correction value on the gage master is zero.
d. Tighten the dial lock screw and remove the gage master.
e. Attach and torque the gage master to the gage once again to verify that the setting is
repeatable. Remove the gage master.
4. Gage the device connector (refer to
a. While holding the gage by the barrel, and without turning the gage or the device, screw
the connecting nut of the device being measured onto the gage, just until you meet
resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting nut to
90 N-cm (8 in-lb). Refer to
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. Read
only
the black
signs;
not
the red
signs.
For maximum accuracy, measure the connector a minimum of three times and take an
average of the readings. After each measurement, rotate the gage a quarter-turn to
reduce measurement variations that result from the gage or the connector face not being
exactly perpendicular to the center axis.
e. Compare the average reading with the observed pin depth limits in
.
Summary of Contents for 85058V
Page 1: ...Keysight Technologies 85058V 1 85 mm Verification Kit User s and Service Guide...
Page 2: ......
Page 3: ...85058V...
Page 7: ...1 General Information...
Page 12: ...1 6 85058V General Information Clarifying the Terminology of a Connector Interface...
Page 13: ...2 Specifications...
Page 20: ...2 8 85058V Specifications Electrical Specifications...
Page 21: ...3 Use Maintenance and Care of the Devices...
Page 36: ...3 16 85058V Use Maintenance and Care of the Devices Handling and Storage...
Page 37: ...4 Performance Verification...
Page 40: ...4 4 85058V Performance Verification Recertification...
Page 41: ...5 Troubleshooting...
Page 43: ...85058V 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart...
Page 47: ...6 Replaceable Parts...
Page 54: ...6 8 85058V Replaceable Parts Replaceable Parts...