3-20 Keysight 85055A User’s and Service Guide
Use, Maintenance, and Care of the Devices
Connections
3-
7.
Remove the protective end cap from the male end of the airline. Align and
insert the male end of the airline center conductor into the female end of
the cable adapter and mate the outer conductors. Connect the nut finger
tight. Do not overtighten.
8.
Using an open-end wrench, hold the sliding female coupling sleeve on the
female end of the outer conductor to keep it from rotating. Refer to
. Use the torque wrench recommended for use with this kit to tighten
the adapter connecting nut to the specified torque. Refer to
Connection Using a Torque Wrench” on page 3-17
9.
Using an open-end wrench to keep the cable adapter from rotating, use
the torque wrench recommended for use with this kit to tighten the airline
male-end connecting nut to the specified torque.
Figure 3-11
Torquing the Connections
How to Separate a Connection
1.
Use an open-end wrench to prevent the device body from turning.
2.
Use the torque wrench to loosen the connector nut.
3.
Complete the disconnection by hand, turning only the connector nut.
4.
Pull the connectors straight apart without twisting, rocking, or bending
either of the connectors.
Do
not
turn the device body. Only turn the connector nut. Damage to the
center conductor can occur if the device body is rotated.
If disconnecting an airline and the airline center conductor does not
disengage from the device center conductor, gently pull the center
conductors apart and then push the airline center conductor back inside
the outer conductor of the airline.
Summary of Contents for 85055A
Page 1: ...User s and Service Guide Type N 50 Verification Kit Keysight Technologies 85055A...
Page 10: ...1 6 Keysight 85055A User s and Service Guide General Information Preventive Maintenance 1...
Page 18: ...2 8 Keysight 85055A User s and Service Guide Specifications Electrical Specifications 2...
Page 58: ...85055 90014 Index Index 6...