3-8 Keysight 85055A User’s and Service Guide
Use, Maintenance, and Care of the Devices
Gaging Connectors
3-
— Initially after every 100 connections, and after that as often as experience
indicates.
Reading the Connector Gage
The gage dial is divided into increments of 0.0001 inch and major divisions of
). For each revolution of the large dial, the smaller
dial indicates a change of 0.01 inch. Use the small dial as the indicator of
multiples of 0.01 inch. In most connector measuring applications, this value
will be zero.
When making a measurement, the gage dial indicator will travel in one of two
directions. If the center conductor is recessed from the zero reference plane,
the indicator will move counterclockwise to indicate the amount of
recession
,
which is read as a negative value. If the center conductor protrudes, the
indicator will move clockwise to indicate the amount of
protrusion
, which is
read as a positive value. Refer to
for definitions of
protrusion and recession.
Figure 3-2
Reading the Connector Gage
Summary of Contents for 85055A