5-2
85038A/F/M
Troubleshooting
Troubleshooting Process
Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow
the steps in
Figure 5-1
.
Figure 5-1 Troubleshooting Flowchart
Clean Devices
Torque 7-16 Devices
to 20 in-lb, 226 N-cm
Load Correct Cal Constants
Recertify Kit
Recalibrate System
Start
(Bad Cal Suspected)
Bad
Measurement?
Devices
Cleaned?
Proper
Torque?
Correct
Cal Constants
Loaded?
Mating
Surfaces Smooth &
Unmarred?
Replace Device(s)
Everything
is Okay
NO
NO
NO
NO
YES
NO
YES
YES
YES
YES
pn43m