2-2
85038A/F/M
Specifications
Environmental Requirements
Environmental Requirements
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a
critical factor in performance. During a measurement calibration, the temperature of the calibration devices
must be stable and within the range shown in
Table 2-1
.
IMPORTANT
Avoid unnecessary handling of the devices during calibration because your fingers are a
heat source.
Table 2-1 Environmental Requirements
Parameter
Limits
Operating temperature
a
a. The temperature range over which the calibration standards maintain conformance to their specifications.
+15
C to +35
C (+59
F to +95
F)
Error-corrected temperature range
b
b. The allowable network analyzer ambient temperature drift during measurement calibration and during
measurements when the network analyzer error correction is turned on. Also, the range over which the
network analyzer maintains its specified performance while correction is turned on.
1
C of measurement calibration temperature
Storage temperature
40
C to +75
C (
40
F to +167
F)
Relative humidity
Type tested, 0% to 95% at 40
C, non-condensing
Summary of Contents for 85038A 7-16
Page 1: ...Keysight Technologies 85038A F M 7 16 Calibration Kits User s and Service Guide...
Page 2: ......
Page 3: ...85038A F M...
Page 6: ...85038A F M...
Page 9: ...1 General Information...
Page 15: ...2 Specifications...
Page 20: ...2 6 85038A F M Specifications Electrical Specifications...
Page 21: ...3 Use Maintenance and Care of the Devices...
Page 29: ...4 Performance Verification...
Page 32: ...4 4 85038A F M Performance Verification Recertification...
Page 33: ...5 Troubleshooting...
Page 38: ...5 6 85038A F M Troubleshooting Contacting Keysight...
Page 39: ...6 Replaceable Parts...
Page 44: ...6 6 85038A F M Replaceable Parts Introduction...
Page 45: ...A Standard Definitions...
Page 50: ...Index 4 85038A F M Index...