background image

Pin

Probes

(Option

001)

The

Pin

Probes

(opt.001)

enables

to

measure

at

or

small

DUT

s

such

as

PC

boards

and

IC

sockets

.

W

arning

Y

ou

must

handle

between

guard

collar

and

socket

(see

gure

below)

to

prevent

an

electrical

shock

when

the

V

output

indicator

on

the

front

panel

of

the

4339B

is

ON.

Note

The

protective

cap

avoids

making

contact

with

adjacent

pins

.

When

measuring

IC

pins

,

attach

the

protective

cap

to

the

tip

of

the

probe

.

Soldering

Sockets

(Option

002)

The

Soldering

Sockets

(opt.002)

helps

make

simple

custom-made

test

leads

.

W

arning

Agilent

T

echnologies

shall

NOT

LIABLE

for

any

damages

or

dangers

to

the

operator

incurred

on

use

of

a

customized

product

except

for

the

16117B

itself.

T

est

xtures

designed

by

users

are

exposed

to

voltages

up

to

1

kV

.

An

operator

may

receive

an

electrical

shock

if

he/she

makes

contact

with

test

xture

components

.

Design

all

components/test

xtures

so

that

an

operator

can

use

the

xture

safely

,

without

being

exposed

to

electrical

shock

hazard.

F

or

example:

Insulate

all

connections

,

solder

joints

,

bare

conductors

.

Design

the

xture

so

that

an

operator

cannot

touch

the

measurement

terminals

when

making

the

measurement.

Provide

warning

labels

to

warn

the

operator

of

the

high-voltage

danger

,

and

to

avoid

touching

any

connections

,

terminals

,

or

DUT

when

the

measurement

is

in

process

or

when

the

high

voltage

is

turned

on.

Operation

3-5

Summary of Contents for 16117B

Page 1: ...Keysight 16117B Low Noise Test Lead Operation and Service Manual...

Page 2: ......

Page 3: ......

Page 4: ......

Page 5: ......

Page 6: ......

Page 7: ......

Page 8: ......

Page 9: ......

Page 10: ...ing charges duties and taxes for products returned to Agilent Technologies from another country Agilent Technologies warrants that its software and rmware designed by Agilent Technologies for use with...

Page 11: ...hether based on contract tort or any other legal theory Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products If you need a...

Page 12: ...ange The manual part number changes when extensive technical changes are incorporated December 1991 First Edition part number 16117 90010 February 1995 Second Edition part number 16117 90020 March 199...

Page 13: ...placement and internal adjustments must be made only by quali ed maintenance personnel Do not replace components with the power cable connected Under certain conditions dangerous voltages may exist ev...

Page 14: ...ion sign denotes a hazard It calls attention to a procedure practice condition or the like which if not correctly performed or adhered to could result in damage to or destruction of part or all of the...

Page 15: ...2 Preparation for Use Introduction 2 1 Initial Inspection 2 1 Ambient Environmental Considerations 2 3 Operating and Storage 2 3 Repackaging the Adapter 2 3 3 Operation Introduction 3 1 Connecting the...

Page 16: ...dapter Features 3 2 3 2 Floating DUT Measurement Option 003 Alligator Clips 3 3 3 3 Grounded DUT Measurement Option 003 Alligator Clips 3 3 Tables 1 1 Furnished Accessories 1 1 1 2 Supplied Options 1...

Page 17: ...2 To operate the 16117B turn to Chapter 3 To order replaceable parts for the 16117B turn to Replaceable Parts in Chapter 4 Product Description The 16117B has been designed to operate speci cally with...

Page 18: ...ainst which the 16117B is tested When shipped from the factory the 16117B meets the speci cations listed in this section Applicable Test Voltage 1000 V maximum Applicable Test Current1 0 5 mA maximum...

Page 19: ...hould be in perfect electrical condition Verify this by carefully performing an incoming inspection to check the adapter for signs of physical damage and missing contents If any discrepancy is found n...

Page 20: ...Description Agilent Part Number Quantity 1 Low Noise Test Lead1 16117B 1 2 Operation and Service Manual2 16117 90060 1 1 Furnished with Alligator Clips if you ordered the 16117B option 003 2 Operation...

Page 21: ...of 0 C to 55 C and relative humidity up to 70 RH at 40 C non condensing The 16117B may be stored within a temperature range of 040 C to 70 and at a relative humidity up to 95 at 40 C non condensing Re...

Page 22: ......

Page 23: ...3 Operation Introduction This chapter describes the features of the 16117B see Figure 3 1 and its connection to the 4339B and DUT Operation 3 1...

Page 24: ...Clip Red Option 003 When making a oating DUT con guration measurement the Red test clip provides a source voltage of up to 1000 V In the grounded DUT measurement con guration the Red test clip is con...

Page 25: ...re 3 3 Warning Do NOT touch the electrode and UNKNOWN connector while the High Voltage indicator is lit which shows the 4339B s output is a high voltage of up to 1000 Vdc maximum You must operate afte...

Page 26: ...117B and 4339B the following items must be carefully observed when performing the OPEN correction Do NOT allow vibration to reach the 16117B when performing an OPEN correction Vibration will result in...

Page 27: ...amages or dangers to the operator incurred on use of a customized product except for the 16117B itself Test xtures designed by users are exposed to voltages up to 1 kV An operator may receive an elect...

Page 28: ...e after turning o the voltage source output and you have con rmed the high voltage indicator is turned o Daily Safety Veri cation Procedure 1 Connect the 16117B to the 4339B 2 Set the source voltage t...

Page 29: ...er than shown in Table 4 1 The listed parts can be ordered from your nearest Agilent Technologies O ce Ordering information should include the Agilent part number and the quantity required Warning The...

Page 30: ...7 61601 1 Cable Assembly Triaxial 2 5 16117 61602 1 Cable Assembly BNC 2 6 16117 61603 1 Cable Assembly Interlock 3 7 16117 25006 2 Black Cover 8 8710 2302 1 Pin Probe Red 4 Option 001 9 8710 2301 1 P...

Page 31: ...This information is subject to change without notice Keysight Technologies 2014 Edition 8 November 2014 16117 90060 16117 90060 www keysight com...

Reviews: