4-8
X, P, and K 11644A
Use, Maintenance, and Care of the Devices
Connections
Tightening a Flange Connection
NOTE
The best connection has symmetrical pressure applied as you gradually tighten the screws.
1. In an “X” pattern (for equal compression), tighten all four screws using a hex ball driver. Do
not
over-tighten. See
2. Visually inspect the connection. Refer to the following section
“Inspecting a Flange Connection”
Figure 4-6 “X” Screw Pattern
Inspecting a Flange Connection
Inspect the flange connection as follows:
1. Place an electric light or white paper behind the connection.
2. Check the flange matings for any gap. A good connection has no gaps between the connected
waveguide flanges, and the waveguide walls are flush. There is no step or offset.
3. Ensure that all four screws are equally tight.
NOTE
The most common cause of measurement error is a poor connection.
Summary of Contents for 11644A K
Page 2: ......
Page 3: ...X P and K 11644A i...
Page 7: ...1 General Information...
Page 12: ...1 6 X P and K 11644A General Information Preventive Maintenance...
Page 13: ...2 Specifications...
Page 19: ...3 User Information...
Page 26: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition...
Page 27: ...4 Use Maintenance and Care of the Devices...
Page 36: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage...
Page 37: ...5 Performance Verification...
Page 52: ...5 16 X P and K 11644A Performance Verification System Operation Checks...
Page 53: ...6 Troubleshooting...
Page 57: ...7 Replaceable Parts...
Page 58: ...7 2 X P and K 11644A Replaceable Parts...
Page 67: ...A Standard Definitions...