X, P, and K 11644A
5-9
Performance Verification
System Operation Checks
The S
21
and S
12
of a Thru Measurement
1. Verify that the calibration is still on.
2. Verify that the averaging factor is set to 512.
3. Connect the standard section to ports 1 and 2 to form a thru connection.
4. To set the display:
a. Press
[S
21
]
.
b.
RESPONSE [REF POSN] [5] [x1]
c. Select
[REF VALUE [0] [x1]
.
d. Press
[SCALE] [.02] [x1]
.
e. Select
MEASUREMENT [RESTART]
.
5. After one complete measurement sweep, the displayed trace should look similar to the typical corrected
trace shown in
6. If necessary, select
Press to Continue
to update the trace.
7. Determine the greatest peak-to-peak deviation (this is any positive peak to any adjacent negative peak).
You can use the markers, as described in the first test, to determine the peak-to-peak value.
8. Record the maximum peak-to-peak value in the appropriate table on
9. Repeat this procedure from step 4, and make an S
12
measurement. Record the measured value in the
.
10. If you used the delta marker mode, don’t forget to turn it off.
Figure 5-5 Typical Corrected S
21
and S
12
Standard Section Ripple
Summary of Contents for 11644A K
Page 2: ......
Page 3: ...X P and K 11644A i...
Page 7: ...1 General Information...
Page 12: ...1 6 X P and K 11644A General Information Preventive Maintenance...
Page 13: ...2 Specifications...
Page 19: ...3 User Information...
Page 26: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition...
Page 27: ...4 Use Maintenance and Care of the Devices...
Page 36: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage...
Page 37: ...5 Performance Verification...
Page 52: ...5 16 X P and K 11644A Performance Verification System Operation Checks...
Page 53: ...6 Troubleshooting...
Page 57: ...7 Replaceable Parts...
Page 58: ...7 2 X P and K 11644A Replaceable Parts...
Page 67: ...A Standard Definitions...