X, P, and K 11644A
5-7
Performance Verification
System Operation Checks
7. To set the display:
a. Press
[S
11
]
.
b. Press
RESPONSE [SCALE] [.1] [x1]
.
c. Select
[REF POSN] [10] [x1]
.
d. Select
[REF VALUE] [0] [x1]
.
e. Press
MEASUREMENT [RESTART]
.
8. After one complete measurement sweep, the displayed trace should look similar to the typical trace
.
9. If necessary, select
Press to Continue
to update the trace.
10. Use the markers to determine the greatest peak-to-peak deviation of the ripple on the displayed trace
(this is any positive peak to any adjacent negative peak):
a. Press
MENUS [MARKER]
and select
delta MODE MENU
.
b. Set the reference to marker 2.
c. Switch between markers 1 and 2 (in the delta mode) and position them to the highest and lowest
peaks. (You may have to practice using the marker features to get the desired results.)
Figure 5-3 Typical S
11
and S
22
Standard Section Ripple
11. Record the maximum peak-to-peak value in the appropriate table on
.
12. Turn off the delta marker mode.
13. Repeat this procedure from step 7, and make an S
22
measurement on the standard section. Record the
measured value in the appropriate table on
Summary of Contents for 11644A K
Page 2: ......
Page 3: ...X P and K 11644A i...
Page 7: ...1 General Information...
Page 12: ...1 6 X P and K 11644A General Information Preventive Maintenance...
Page 13: ...2 Specifications...
Page 19: ...3 User Information...
Page 26: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition...
Page 27: ...4 Use Maintenance and Care of the Devices...
Page 36: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage...
Page 37: ...5 Performance Verification...
Page 52: ...5 16 X P and K 11644A Performance Verification System Operation Checks...
Page 53: ...6 Troubleshooting...
Page 57: ...7 Replaceable Parts...
Page 58: ...7 2 X P and K 11644A Replaceable Parts...
Page 67: ...A Standard Definitions...