TROUBLESHOOTING GUIDE
IF YOU HAVE ANY OTHER PROBLEMS, PLEASE CALL THE FACTORY.
PROBLEM
POSSIBLE CAUSES
SOLUTIONS
Power is applied to unit but
the display does not light.
Unit works but occasionally
the display freezes or skips
counts.
Input signal is connected but
the unit does not work.
Display reading is inaccurate.
1. AC or DC power wiring is
incorrect.
1. Line noise is effecting the
processor due to a current
spike or surge.
1. Input wiring is incorrect
2. High and low scaling set-
tings are incorrect.
3. Transmitting device is de-
fective.
4. Unit is defective.
1. Input wiring is incorrect.
1. Recheck power wiring.
1. Use a different power sup-
ply or install a surge suppres-
sor.
1. Recheck input wiring.
2. Recheck high and low scal-
ing settings.
3. Replace transmitting device.
4.To confirm set meter for 0-
10V input, low @ 0; high @
10. Apply a 0-10V signal to
the voltage input (pin 5). When
running, the meter should dis-
play the voltage value that is
applied. If not call factory for
an RMA#.
1. Be sure that voltage signals
are connected to voltage in-
put (pin 5) and current signals
are connected to current input
(pin 4).