9 Error messages
Error
Possible cause
Remedy
Err01
Signal of the weighing cell unstable
Automatic
Err02
Balance overloaded
Automatic after unloading
Err03
Gross negative. This action is not allowed
Automatic
Err04
Outside the zero range
Press any button
Err05
Exactness of piece calculation insufficient
Press any button
Err06
Load cell signal too high
Automatically to correction
and input
Err07
Load cell signal too low
Automatically to correction
and input
Err08
Calibration outside the range (negative)
Automatic
Err09
Calibration outside the range (signal too low)
Automatic
Err10
Calibration value of the 2nd (3rd) point lower
than the value of the 1st (2nd) point
Automatic
Err11
Calibration from piece counting mode
Automatic
Err97
Calibration blocked
Automatic
Err98
The calibration value must be higher than the
previous one
Automatic
Err99
Actuation only allowed on the display device
Automatic
VHT-BA-e-0910